National Instruments Corporation
5
NI TB-2706 Installation Guide
6.
Tighten or replace the strain-relief screws.
7.
Replace the terminal block cover by sliding the indentations on the cover under the tabs on the base.
Tighten the top cover screw.
Installing the Terminal Block
Refer to Figure 4 as you complete the following steps to connect the terminal block to the PXI/PXIe
module I/O connector.
Figure 4.
Connecting the TB-2706 to an M Series or PXI-6143 Device
Caution
The connectors of both the PXI/PXIe module and the terminal block are polarized. You can
attach them in only one way. Do
not
force the terminal block when inserting it into or removing it
from the PXI/PXIe module I/O connector. Make sure that the chassis is powered off before inserting
the PXI/PXIe module.
1.
Install the PXI/PXIe module into the chassis and tighten the two module screws. You must install
the TB-2706 terminal block on the PXI/PXIe module
after
the module is installed in the chassis.
2.
Guide the terminal block onto the connector.
3.
Tighten the two terminal block mounting screws.
Caution
The TB-2706 is not designed for input voltages greater than 42 V
pk
/60 VDC, even if you
install a voltage divider that reduces the voltage to within the input range of the DAQ device. Input
voltages greater than 42 V
pk
/60 VDC can damage the TB-2706, any device connected to it, and the
host computer. Overvoltage also can cause an electric shock hazard for the operator. National
Instruments is
not
liable for damage or injury resulting from such misuse.
1
Terminal Block Mounting Screws
2
PXI/PXIe Chassis
3
PXI/PXIe Module Connector