Chapter 3 Operating the VXI-1200
© National Instruments Corporation
3-11
VXI-1200 User Manual
Basic Functional Check
The functional check consists of checking the VXI-1200 power supply
voltages to the VXIbus limits at the rear-panel connector J415 using a
digital voltmeter.
Referring to Table 3-5, connect one lead of the voltmeter to a supply
pin at J415. Connect the reference lead of the voltmeter to one of the
ground pins. Compare each voltage reading to the values listed in
Table 3-5.
Table 3-5. Power Supply Voltages at J415
J415 Pin
Supply
VXI Acceptable
Voltage Range
3
+5 V
4.875 to 5.25 V
10
-12 V
-12.6 to 11.64 V
1
-24 V
-25.2 to -23.28 V
12
-2 V
-2.1 to -1.9 V
2
+12 V
11.64 to 12.6 V
9
+24 V
23.28 to 25.2 V
4
-5.2 V
-5.46 to -5.044 V
7,11
Logic Ground
—
If the voltages are within the specified ranges, the mainframe complies
with VXI voltage limit specifications.