Appendix
A Specifications
VXI-1200 User Manual
A-6
© National Instruments Corporation
Environmental
Characteristic
Description
Temperature
Operating
Non-operating
Meets the limits stated in MIL-T-28800E for
Type III, Class 5
0
o
to 55
o
C
-40
o
to 71
o
C
Relative Humidity (RH)
Operating
Non-operating
95
±
5% RH noncondensing
75
±
5% RH above 30
o
C
45
±
5% RH above 40
o
C
95
±
5% RH at <55
o
C
Altitude
Operating
Non-operating
4,570 m (10,000 ft.)
12,190 m (15,000 ft.)
Vibration
0.013 in. double amplitude, 5 to 55 Hz
Functional Shock (Operating)
Half sine, 30 g, 11 ms duration.
EMI
FCC Class A Verified, EC Certified
Immunity, Enclosure, Radio
Frequency Electromagnetic Field
No mainframe failures when the mainframe is
subjected to a 3 V/m electromagnetic field over the
frequency range of 27 to 500 MHz
Immunity, Enclosure,
Electrostatic Discharge (ESD)
Up to 8 kV with no change to control settings or
impairment of normal operation
Immunity, Fast Transients,
Common Mode
Port Peak Voltage Tr/Th Rep Freq.
Signal&Control 0.5 kV 5/50 ns 5 kHz
DC Power 0.5 kV 5/50 ns 5 kHz
AC Power 1.0 kV 5/50 ns 5 kHz