Chapter 2 Installation and Configuration
© National Instruments Corporation
2-7
VXI-1500 User Manual
Installing VXI Modules
Caution:
Turn off the mainframe power before installing VXI modules.
Install C-size modules directly into the 13 C-size slots of the mainframe
by first placing the module’s card edges into the front module guides (top
and bottom). Slide the module to the rear of the mainframe until the
module connectors mate solidly with the backplane connectors. Secure
the module’s front panel to the mainframe using the module’s front-panel
mounting screws.
Installing Filler Panels
To optimize module cooling performance, you should install filler panels
into unused or empty slots. Secure with the captive mounting screws.
Your kit includes an assortment of filler panels of various widths. Use the
widest filler panels first to minimize the number of panels that you need
to install. To further improve cooling performance, you can install
airflow blockers into unused slots.
Routing Cables
The VXI-1500 has a built-in cable channel providing a variety of options
for instrument cable routing. The channel is located in the bottom section
of the mainframe. Cables may be routed out the front by removing the
cable access panel via the four thumbscrews. See Figure 1-2 for a
diagram of the cable access panel. Likewise, cables can be routed out the
rear by removing the rear cable access panel. See Figure 1-3 for a
diagram of the rear cable access panel. Cable apertures are also located in
the bottom of the chassis.
Note:
Use the extra grommeting included in your kit to help protect your cables
from possible damage by the front or rear aperture edges.