VXI-1501 Advanced Rack Mount Kit

2

ni.com

Installing the Rack Mount Flanges and Handles

Caution

Use both handles when lifting units that weigh more than 40 lbs. (18.2 kg).

Do

not

lift using the handles on units weighing more than 80 lbs. (36.5 kg). Do

not

mount

the chassis using only the flanges. The chassis must be supported by this slide kit, the rail
kit (offered by National Instruments as the Basic Rack Mount Kit), or a cabinet shelf.

Follow these steps to install the rack mount flanges and handles:

1.

Remove the side trim strips from the VXI-1501, as shown in Figure 1.

Figure 1.

Installing the Rack Mount Flanges and Handles

2.

Attach the rack mount flanges and front handle assemblies with three
M4

×

16L screws per side (included with the kit). See Figure 1.

1

Handle (Qty. 2)

2

Rack Mount Flange (Qty. 2)

3

Side Trim Strip

3

1

2

Summary of Contents for VXI-1501

Page 1: ...t a cable tray kit a horizontal filler panel flanges for mounting the chassis to the cabinet and handles for easy handling of the chassis when sliding the chassis into and out of the VXI cabinet For installation instructions see the Installing the Rack Mount Flanges and Handles section You must install all accessories in the order shown in this installation guide with the exception of the horizont...

Page 2: ... be supported by this slide kit the rail kit offered by National Instruments as the Basic Rack Mount Kit or a cabinet shelf Follow these steps to install the rack mount flanges and handles 1 Remove the side trim strips from the VXI 1501 as shown in Figure 1 Figure 1 Installing the Rack Mount Flanges and Handles 2 Attach the rack mount flanges and front handle assemblies with three M4 16L screws pe...

Page 3: ...ng the Cable Tray Follow these steps to install the cable tray 1 Place the VXI 1501 on its rear feet 2 Mount the cable tray using three of the M5 10L flathead screws packaged with the cable tray spacers on each side of the chassis Using the tray holes as a guide make sure the tray is oriented correctly as shown in Figure 2 1 VXI 1501 Chassis 2 Cable Tray 3 Standard Tray Depth Position 44 5 mm 1 75...

Page 4: ... rack slide Note that the left and right slide rails are not identical and thus are labeled RH for the right rail and LH for the left rail When separating the inner rails note which one goes with the right slide and which goes with the left slide 1 Flange 2 Spacer 3 Pan Head Screw M5 12L Qty 8 Two Per Bracket 4 Slide Label 5 Adapter Bracket Qty 4 6 Rear Pan Head Screw M5 12L Qty 4 Two Places Per R...

Page 5: ...12L pan head screws and two M5 hex nuts per bracket included with the adapter kit Some adjustments may be necessary depending on your particular rack Tighten the nuts after making any adjustments Note Depending on your particular VXI rack you may be able to mount the slide rack assembly directly to the cabinet rails without using the adapter kit Should you try this remember that depending on which...

Page 6: ...sis to the front rack rails with the hardware included in the handle and flange kit Installing the Horizontal Filler Panel Once the VXI 1501 is mounted in the rack you may want to install the 1U filler panel above it as shown in Figure 4 This panel allows adequate air ventilation for the vent holes on top of the chassis should another piece of equipment be installed above it Figure 4 Installing th...

Reviews: