Appendix A
VXI-1501 Specifications
© National Instruments Corporation
A-3
VXI Ripple/Noise
1
Maximum Power Output
Uncertainty in shutdown is due to variable loading. Heavy loading on
+5 V, –2 V, and –5.2 V will cause shutdown at the lower edge of the band.
Heavy loading of the +12 V, –12 V, +24 V, or –24 V will cause shutdown
at the higher edge of the band. Low line voltage (90 VAC) will cause
shutdown at the lower edge of the band.
Humidity ................................................ 10-90% RH noncondensing
Temperature range
Nonoperating .................................. –20 °C to +70 °C
Operating ........................................ 0 °C to 50 °C
Shock ..................................................... 30 g, 11 ms, half sine
3 shocks per face
Random vibration
Nonoperating .................................. 5-500 Hz, 2.46 g
rms
Operating ........................................ 5-500 Hz, 0.31 g
rms
1
Per VXI Specification rev. 1.4.
DC Output
Voltage
Allowed
Variation
Max. DC Load
Ripple/Noise
Max. Induced
Ripple/Noise
+5 V
+0.25/–0.125
50 mV
50 mV
+12 V
+0.06/–0.36
50 mV
50 mV
–12 V
–0.60/+0.36
50 mV
50 mV
+24 V
+1.2/–0.72
150 mV
150 mV
–24 V
–1.2/+0.72
150 mV
150 mV
–5.2 V
–0.26/+0.156
50 mV
50 mV
–2 V
–0.10/+0.10
50 mV
50 mV