TROUBLESHOOTING 

 

PROBLEM 

REASON FOR PROBLEM 

SOLUTION 

Light fails to operate. 

Outlet inoperative. 

Have qualified service technician 
repair outlet. 

 

 

 

 

AC power cord not connected. 

Plug into outlet. 

 

 

 

 

Lamp burned out. 

Replace lamp. 

 

 

 

(131/132-CLED only)

 

LED bulb burned out. 

Replace LED. 

 

 

 

(131/132-CLED only)

 

Fuse blown. 

Replace fuse. 

Image does not remain in focus 
 

Stage of microscope drops from 
its own weight. 

Adjust tension control. 
 

Image will not focus 

Rack stop not set at proper 
position. 

Adjust rack stop. 

 

 

 

 

Slide upside down. 

Place slide on stage with cover 
slip up. 

 

 

 

 

Slide cover slip too thick. 

Use 0.17mm thick cover slip  
     (No.1 cover slip) 

Poor resolution 
     (image not sharp) 

Objective lenses dirty. 

Clean objective lenses. 

 

 

 

 

Eyepiece lens dirty. 

Clean eyepiece lenses. 

 

 

 

 

Too much light. 

Adjust disc diaphragm. 

Spots in field of view. 

Eyepiece or condenser lens dirty.  Clean lens. *** 

 

 

 

 

Specimen slide dirty. 

Clean slide. 

***Spots in field of view can also result from dirt on inside of eyepiece.  It is recommended that you have 
service technician clean inside of lens. 

 
 

OPTIONAL ACCESSORIES AND PARTS: 

 
#610-045R 

WF10x eyepiece w/reticle, 10mm/100div. 

#615-045 

WF15x eyepiece w/pointer 

#760-045 

DIN 60xR objective lens, N.A. 0.90 

#800-001 

Replacement LED bulb 

(131-CLED & 132-CLED only) 

#800-102 

Replacement bulb, 115v 20 watt tungsten D.C. medium base 

(131 & 132 only)

 

#801-160 

Replacement fuse, 0.5 amp, time delay 

(131-CLED & 132-CLED only) 

#910 

 

Graduated mechanical stage, side mount adjustments (installed by end user) 

#951 

 

Dustcover, 16” tall x 13”, heavy vinyl with stitched seams. 

#975-001 

Carrying case, anodized aluminum, fabric lining, accessory pockets, Velcro straps, keyed lock.  Fits all 
models in series, except dual head models. 

#975-002 

Carrying case, anodized aluminum, fabric lining, accessory pockets, Velcro straps, keyed lock.  Fits dual 
head models. 

#975-138 

Carrying case, heavy plastic w/foam liner.  Fits all models in series, except dual head models. 

 

LIMITED LIFETIME WARRANTY 

 
Please see our website, 

www.nationaloptical.com

, for complete warranty details and exclusions. 

 
 

 

 

 

 

 
 
 
 
 

(Revised 6/24/14) 

Summary of Contents for 131

Page 1: ... These models share common features nomenclature and method of operation The only difference are Models 131 132 feature electric illumination from a tungsten bulb Models 131 CLED 132 CLED feature electric illumination from a LED bulb Models 132 132 CLED have both 30 inclined and vertical eyepiece tubes each supplied with WF10x eyepiece Copyright 1 2 01 National Optical Scientific Instrument Inc ...

Page 2: ...STAGE CLIPS Two locked on clips hold specimen slide in place on stage Note Your microscope is already drilled and tapped to accept an optional mechanical stage Mechanical stage replaces stage clips and permits precise mechanical manipulation of the specimen slide 6 STAGE Platform of the microscope where the specimen slide is placed 7 CONDENSER LENS An 0 65 N A condenser lens fixed in center of sta...

Page 3: ...therwise you will not be able to see anything when looking through the microscope b Standard lenses provided with your microscope are a widefield 10x eyepiece two eyepieces supplied with models 132 and 132 CLED 4x 10x and 40x objectives The 40x objective lens has a special spring retractable mechanism which retracts slightly if front of lens comes in contact with the specimen slide See chart below...

Page 4: ... swab or lens tissue To clean objective lenses do not remove objectives from microscope Clean front lens element only following same procedure NOTE Fingerprints or other matter on the front lens element of the objective lens is the single most common reason that you will have difficulty in focusing the microscope Before having costly servicing done or before returning to National for warranty repa...

Page 5: ...tly Replace with a 110v 20 watt lamp with double contact medium bayonet base This is a standard commonly used replacement bulb National 800 102 Carefully wipe lamp to insure it is clean and free of all fingerprints Replace field lens housing assembly b Changing LED bulb on Models 131 CLED 132 CLED To open the illuminators field lens housing use 0 9mm L hex key wrench supplied with your microscope ...

Page 6: ...n also result from dirt on inside of eyepiece It is recommended that you have service technician clean inside of lens OPTIONAL ACCESSORIES AND PARTS 610 045R WF10x eyepiece w reticle 10mm 100div 615 045 WF15x eyepiece w pointer 760 045 DIN 60xR objective lens N A 0 90 800 001 Replacement LED bulb 131 CLED 132 CLED only 800 102 Replacement bulb 115v 20 watt tungsten D C medium base 131 132 only 801...

Reviews: