disconnect the charger from the AC wall outlet and the output cord from recharging socket after 12 hours.  
Batteries and charger may feel warm when charging, and unplugging the recharger is a safety precaution. 
 
You may operate the microscope light even while it is being recharged.  Simply flip light switch to “on” position and 
continue using microscope while the recharger is fully engaged. 

 

b.  Replacing batteries: 

 

Your microscope includes 3 rechargeable AA nickel metal hydride batteries.  These may be recharged up to 500 
times, but if you observe that a recharge is providing significantly less than 40 hours of operation.  It is probably 
time to replace to batteries. 

 

IMPORTANT WARNING:  DO NOT USE REGULAR ALKALINE BATTERIES IN THIS MICROSCOPE.  ANY 
ATTEMPT TO RECHARGE ALKALINE TYPE BATTERIES COULD RESULT IN BATTERIES EXPLODING. 

 

Gently lay microscope on its side or back.  Observe small door in center of microscope bottom base.  Loosen 
slotted screw on door and open.  Battery case is mounted on inside of door.  Using small Phillips screwdriver, 
carefully remove Phillips screw that holds battery case together.  Slide lid of case straight out to remove and 
expose batteries.  Remove all 3 batteries and replace with new rechargeable AA nickel metal hydride batteries, 
making certain to insert with correct polarity according to markings on battery holder.  Replace lid, close and 
secure door. 

 

Follow instructions on new battery packaging to determine if they are already charged, or if they should be charged 
before initial use.  If recharging is required, following directions in “3.a” above. 

 

c.  Replacing LED element: 

 

An LED “bulb” will last up to 50,000 hours, so you don’t have to do this exercise very often. 

 

To open the illuminators field lens housing, use 0.9mm “L” type hex key wrench supplied with your microscope.  
Loosen hex screws on lens housing.  Remove lens housing to expose LED “bulb”.  Remove bulb by grasping the 
plastic base of bulb and gently pulling straight up.  Insert new LED “bulb”, replace lens housing and tighten hex 
screw to secure lens housing in place. 

 
TROUBLESHOOTING 

 

PROBLEM 

REASON FOR PROBLEM 

SOLUTION 

Light fails to operate. 

Batteries fully discharged. 

Recharge batteries. 

 

 

 

 

Rheostat control not turned far 
enough. 

Turn rheostat to increase light 
intensity. 

 

 

 

 

Light switch in ‘off’ position. 
 
LED “bulb” burned out. 
 
Fuse Blown 

Turn light switch on. 
 
Replace LED “bulb”. 
 
Replace Fuse 

Image does not remain in focus 
 

Stage of microscope drops from 
its own weight. 

Adjust tension control. 
 

Image will not focus 

Rack stop not set at proper 
position. 

Adjust rack stop. 

 

 

 

 

Slide upside down. 

Place slide on stage with cover 
slip up. 

 

 

 

 

Slide cover slip too thick. 

Use 0.17mm thick cover slip  
     (No.1 cover slip) 

Poor resolution 
     (image not sharp) 

Objective lenses dirty. 

Clean objective lenses. 

 

Eyepiece lens dirty. 

Clean eyepiece lenses. 

 

 

 

 

Too much light. 

Adjust diaphragm. 

Spots in field of view. 

Eyepiece or condenser lens dirty. 

Clean lens. *** 

 

 

 

 

Specimen slide dirty. 

Clean slide. 

***Spots in field of view can also result from dirt on inside of eyepiece.  It is recommended that you have 
service technician clean inside of lens. 

Summary of Contents for 211

Page 1: ...08 Tri County Parkway Schertz Texas 78154 Phone 210 590 9010 Fax 210 590 1104 INSTRUCTIONS FOR MODELS 211 212 Corded MODELS 211 RLED 212 RLED Rechargeable Cordless COMPOUND MICROSCOPES Copyright 1 14 2013 National Optical Scientific Instrument Inc ...

Page 2: ...e Objective lens Specimen holder mechanical stage Iris diaphragm lever Condenser lens Stage Illuminator field lens housing Base Coarse focus knob Fine focus knob Rheostat control Tension adjustment On Off switch and fuse holder back of base Swing out filter holder Rack stop set screw ...

Page 3: ...ecimen 4 MECHANICAL STAGE Permits precise mechanical manipulation of the specimen slide 5 STAGE Platform of the microscope where the specimen slide is placed 6 CONDENSER LENS 1 25 N A Abbe condenser lens positioned under center of stage condenses light rays from substage illumination and fills the back lens element of objective lens to improve image resolution 7 IRIS DIAPHRAGM Iris opening is cont...

Page 4: ...ger is also equipped with an automatic trickle charge feature the red LED indicator lamp located on recharger will be illuminated when batteries are receiving maximum charge After batteries are charged the red LED indicator lamp will turn to green and charger automatically switches to trickle charge The charger can be left plugged in but for safety reasons it is a good idea to disconnect the charg...

Page 5: ...kes the specimen slide With fine focus adjustment at mid range the rack stop has been adjusted at the factory to assure the 40x and 100x lens will clear the thickness of a normal specimen slide and cover slip However if the rack stop has been improperly adjusted or if using a thicker than normal slide or cover slip moving the 40x and 100x lenses too quickly or carelessly could cause damage to the ...

Page 6: ... clockwise to loosen tension Use of a wide rubber band will provide a better grip on the tension adjustment collar After adjusting tighten the set screw to lock collar in place NOTE It is recommended that you leave the tension as loose as possible for ease of focusing yet not so loose that it permits the stage of microscope to drift downward from its own weight and cause the microscope to drift ou...

Page 7: ...rmine if they are already charged or if they should be charged before initial use If recharging is required following directions in 3 a above c Replacing LED element An LED bulb will last up to 50 000 hours so you don t have to do this exercise very often To open the illuminators field lens housing use 0 9mm L type hex key wrench supplied with your microscope Loosen hex screws on lens housing Remo...

Page 8: ...for rechargeable LED microscopes 801 160 Replacement fuse 0 5 amp time delay 951 Dustcover 16 tall x 13 heavy vinyl with stitched seams 975 001 Carrying case anodized aluminum fabric lining accessory pockets Velcro straps keyed lock Fits all models in series except dual head models LIMITED LIFETIME WARRANTY Please see our website www nationaloptical com for complete warranty details and exclusions...

Reviews: