ALGO® 5 Newborn Hearing Screener Service Manual
P/N 000976 Rev.A
3.3. Troubleshooting
the
Isolation Transformer
Follow the steps below to change the Isolation Transformer’s fuse:
•
Unplug the power cable leading from the Isolation Transformer to the wall.
•
Locate the power inlet module on the Isolation Transformer. Note the orientation of
the power inlet module so as to put it back in the same direction.
•
Using a flat head screwdriver, open the power inlet module to access the fuse holder
(refer to FIGURE 21: Open the power inlet module ).
FIGURE 21: Open the power inlet module
•
Using a flat head screwdriver, remove the fuse holder from the drawer (refer to
FIGURE 22)
FIGURE 22: Access to the fuse holder