Natus ErgoJust and ErgoJust ICU Cart System
July 6, 2020
3-9
4. Remove the video
pole
-
Natus ErgoJust
LTM and ErgoJust
ICU
IMPORTANT
: Do NOT remove the mounting screws during the following step. If the
mounting screws are removed, the mounting plate(s) will slide down. Ensure the plates are
orientated with the
convex
side of the plate pointing inward toward the bottom mast (Figure 1)
when remounting the video pole.
Figure 1: Proper mounting plates orientation.
8. Loosen the four video pole mounting screws sufficiently to allow the video pole to be lifted
away from the bottom mast (Figure 2).
9. Using an assistant, carefully lift the video pole away from the bottom mast and set it aside
(Figure 3).
Legend Description
A
Concave side facing outward
B
Bottom mast
Figure 2: Loosen the four video pole
mounting screws.
Figure 3: Lift the video pole away from the
bottom mast.
Legend Description
A
Video pole
B
Loosen
C
Bottom mast