8 Testing with Zodiac Quick Check - Stand-Alone
40
MADSEN Zodiac
Probe color
Status
Red
•
The right test ear has been selected
•
The device is in idle mode
Blue
•
The left test ear has been selected
•
The device is in idle mode
Green
•
The test is running
Yellow
•
Leak
Otosuite indications
Color
Status
Green
•
During measurements, Otosuite shows a green background to the online
values.
Probe leakage
Stand-alone
If there is a probe leak during testing, this icon appears on the Madsen Zodiac and Otosuite screens.
PC-based
If there is a probe leak during testing, this icon appears on the Otosuite screen.
If you are using the Quick Check probe, the leakage indication will remain until you remove the probe
from the ear and try again.
Probe blocked
Stand-alone
If the probe is blocked during testing, this will be shown on the Madsen Zodiac and Otosuite screens.
PC-based
If the probe is blocked during testing, this will be shown on the Otosuite screen.
If you are using the Quick Check probe, the blocked probe indication will remain until you remove the
probe from the ear and try again.
8.2
The Quick Check measurement screen
When the measurement starts, you will see the measurement being performed real-time on the screen.