9 Testing with Zodiac Diagnostic or Clinical - Stand-Alone
45
MADSEN Zodiac
6.
A flat tympanogram together with an abnormally
small ear canal volume (ECV) indicates that the
probe is blocked
7.
Any leakage will interrupt the test. The probe will
indicate if there is a leak.
When you get a leak detection signal, you must
briefly remove the probe from the ear before
trying again.
8.
Make sure that the eartip fits well. This will
minimize the risk of blocking the probe tip against
the ear canal wall. A flat tympanogram together
with an abnormally small ear canal volume (ECV)
indicates that the probe is blocked.
A measurement will not auto start if the ear
canal volume reading is less than 0.1.
9.
Any leakage will interrupt the test. The probe will
indicate if there is a leak.
Probe status and leakage
The light indicators
Light indicators light up in color in the probe to indicate different states. The control pad on the shoulder strap always
shows the selected ear color if you are using the diagnostic probe.
Probe color
Status
Red
•
The right test ear has been selected
•
The device is in idle mode
Blue
•
The left test ear has been selected
•
The device is in idle mode
Green
•
The test is running
Yellow
•
Leak
Otosuite indications
Color
Status
Green
•
During measurements, Otosuite shows a green background to the online
values.