6.   To track treatment time, reset timer to zero.

 Press and hold     

        Treatment Timer Reset Button for approximately two seconds. 

7.   Monitor the patient during treatment. 

8.   Turn on white exam light as needed. 

Press        exam light  

button on. 

9.   Turn off white exam light when finished.

 Press        exam 

light button off, or press        On/Standby button to resume blue 
treatment light. 

10.  Turn off blue treatment light when finished.

  

Press        On/Stand-by button.

Summary of Contents for neoBLUE Compact

Page 1: ... Quick Guide The neoBLUE compact LED Phototherapy System is intended for the treatment of neonatal hyperbilirubinemia The light can be used for infants in a bassinet incubator open bed or radiant warmer ...

Page 2: ...and 15µW cm2 nm at the Low setting at a distance of 35 cm 13 75 in from the baby If adjusted record your intensity settings below Note This neoBLUE device has been adjusted to read the following light intensity levels at the following distances High _________ µW cm nm at _________cm or in Low _________ µW cm nm at _________cm or in ...

Page 3: ...ction Indicator Treatment Timer Display High Intensity Setting Button High Intensity Indicator White Exam Light Button Service Indicator Treatment Timer Reset Button Low Intensity Indicator On Standby Button Low Intensity Setting Button ...

Page 4: ...ototherapy 3 Position light over infant The light can be placed on top of an incubator or used with the arm and roll stand for positioning the light with a bassinet open bed radiant warmer or incubator 4 Turn on blue treatment light Press On Stand by button on the control panel 5 Select High or Low intensity Press High or Low intensity setting buttons as appropriate for the patient ...

Page 5: ...wo seconds 7 Monitor the patient during treatment 8 Turn on white exam light as needed Press exam light button on 9 Turn off white exam light when finished Press exam light button off or press On Standby button to resume blue treatment light 10 Turn off blue treatment light when finished Press On Stand by button ...

Page 6: ...e must be taken to ensure a safe operating environment Secure power cord to minimize risk of tripping Warning Maintain a minimum treatment distance of 35 cm 13 75 in between the light and infant for optimal uniformity Warning Select only infants for whom phototherapy has been prescribed Warning The intensity level and duration of treatment should be prescribed by the physician for each patient War...

Page 7: ...erify that the baby s eyes are protected and free of infection Warning Regular monitoring during treatment is recommended Caution Federal Law U S restricts this device to sale or use by or on the order of a physician or properly licensed practitioner For a complete list of warnings cautions and additional information please refer to the neoBLUE compact user manual For service information please re...

Page 8: ...A 98108 USA 1 800 303 0306 1 650 802 0400 www natus com Natus Europe GmbH Robert Koch Str 1 82152 Planegg Germany P N 022093 Revision Date 2017 03 24 Maisels MJ Watchko JF Bhutani VK Stevenson DK An approach to the management of hyperbilirubinemia in the preterm infant less than 35 weeks of gestation Journal of Perinatology 2012 32 660 664 ...

Reviews: