Maximum Current Limits are listed below for each test and standard.
TABLE 2.1
Power Cord Ground Integrity
100 milliohms or less
TABLE 2.2
Earth Leakage Current
AC Line Cord
Polarity
Line Cord
Neutral Cord
IEC 60601-1
UL 2601-1 ANSI/NFPA
99
Normal
Closed
Closed
500 µA
250 µA
Reversed Closed Closed 500
A
250 µA
Normal
Open
Closed
1000 µA
500 µA
Normal
Closed
Open
1000 µA
500 µA
TABLE 2.3 Main Enclosure Leakage Current
AC Line Cord
Polarity
AC Line
Cord
Neutral
Conductor
AC Line Cord
Ground
Conductor
IEC 60601-1
UL 2601-1 AMI/ANSI-
E1
Normal
Closed
Closed
100 uA
300 uA
Normal
Closed
Open
500 uA
300 uA
Normal
Open
Closed
100 uA
300 uA*
Reversed
Closed
Closed
500 uA
300 uA
Reversed
Open
Closed
500 uA
300 uA*
Reversed
Closed
Open
500 uA
300 uA
* AAMI/ANSI-ES1 does not include opening the neutral line conductor
neoBLUE® cozy Service Manual
9
P/N 051915 Rev.
D