background image

 

SYMPTOM 

PROBABLE CAUSE 

CORRECTIVE ACTION 

Air vents are 
blocked 

Verify that vents are not blocked. 
If the patient surface is above 40°C (104°F), 
verify that the fan is functioning correctly.   

Device not used 
within specified 
operating 
temperature 

Operating conditions for the neoBLUE cozy 
light are 20 – 30°C (68 - 86°F).  Verify that 
temperature is within range, and adjust 
ambient temperature or move the device to 
a cooler area. 

Excessive heat is 
being trapped 
underneath the 
device 

Remove any mattress or cushioning from 
underneath the device so that the device is 
operating on a hard, flat surface. 

If none of the above corrects the problem, open the main 
enclosure and check for the following:

 

MATTRESS SURFACE 
HAS BECOME 
EXCESSIVELY HOT 
(ABOVE 40°C/104°F) 

Fan is not running 

The fan is not a user serviceable 
component.  Replace LED PCB.  Contact 
Natus Technical Service or authorized 
service provider. 

Air vents are 
blocked 

Verify that vents are not blocked. 
If the infant becomes too warm, verify that 
the fan is functioning correctly.   To reset 
Indicator LED, turn device off and on again. 

Device not used 
within specified 
operating 
temperature 

Make sure that the device is not used in 
ambient temperature greater than 30

°

C, 

causing device to turn off to prevent 
overheating.  To reset Indicator LED, turn 
device off and on again. 

Device placed on a 
soft surface (i.e. 
mattress, bed) 

The device should be placed on a hard 
surface during operation.  The mattress or 
any soft surface in contact with the center 
bottom of the device will trap heat, causing 
the device to overheat, triggering the 
indicator LED. 

Filter has become 
clogged 

Clean filter per Section 2.1.2. 

Main enclosure fan 
is not running 

Verify that fan is able to spin freely.  If 
necessary, replace LED PCB (fan is not 
available separately).  Contact Natus 
Technical Service or authorized service 
provider. 

INDICATOR LED 
TRIGGERS 
(TEMPERATURE 
CUTOFF POINT 
REACHED) 
 

Component on 
PCB board has 
become faulty.  

Replace LED PCB (see Chapter 4).  Contact 
Natus Technical Service or authorized 
service provider. 

neoBLUE® cozy Service Manual 

12 

P/N 051915 Rev.

 

D

 

Summary of Contents for neoBLUE cozy

Page 1: ...SERVICE MANUAL neoBLUE cozy Service Manual i P N 051915 Rev D...

Page 2: ...rt without written consent from Natus Medical Incorporated The content of this manual may change without notice Natus Medical Incorporated 5900 First Avenue South Seattle WA 98108 USA Telephone 1 650...

Page 3: ...ifying General Performance 6 2 2 2 Adjusting Light Intensity 7 2 3 Safety Testing 8 3 Troubleshooting the neoBLUE cozy light 10 4 Repair Procedures 15 4 1 Removing the Plastic Cover 15 4 2 Removing th...

Page 4: ...W cm2 nm Blue LED components emit light in the range of 400 550 nm peak wavelength 450 470 nm This range corresponds to the spectral absorption of light by bilirubin and is thus considered to be the...

Page 5: ...ent to the side of a bassinet or railing A channel on the bottom of the light enclosure provides for safe routing of the power cord 1 4 Additional References Note neoBLUE cozy device User Manual is av...

Page 6: ...nsult Instructions for Use Always protect the baby s eyes with eye patches or equivalent Single use only Do not use in presence of enriched oxygen or flammable gases Ambient Operating Temperature DC V...

Page 7: ...the Exterior of the Light Enclosure and Mattress Section 2 1 1 Cleaning the Filter Section 2 1 2 Cleaning the Interior of the Light Enclosure Section 2 1 3 Warning Disconnect UUT from AC power before...

Page 8: ...he housing An extra filter is provided so the UUT can be used while a filter is being cleaned FIGURE 1 REMOVING THE FILTER HOUSING Place the filter back into the housing ensuring it completely covers...

Page 9: ...before adjusting light intensity Use of a radiometer with a HOLD feature is necessary if the radiometer display cannot be read while the radiometer sensor is faced down Intensity measurements are bas...

Page 10: ...f all previous steps generate satisfactory results the UUT performance has been verified and the UUT can be returned to the user Switch the UUT to stand by and return it to service 2 2 2 Adjusting Lig...

Page 11: ...he UUT to service 6 If steps 1 4 do not comply refer to Chapter 3 Troubleshooting the neoBLUE cozy light 2 3 Safety Testing The safety tests listed below meet the standards of and should be performed...

Page 12: ...1000 A 500 A Normal Closed Open 1000 A 500 A TABLE 2 3 Main Enclosure Leakage Current AC Line Cord Polarity AC Line Cord Neutral Conductor AC Line Cord Ground Conductor IEC 60601 1 UL 2601 1 AMI ANSI...

Page 13: ...ning LED turns on blue light goes off LED string s goes out Multiple LED strings go out Cannot attain appropriate intensity Light does not maintain intensity Excess noise When opening the neoBLUE cozy...

Page 14: ...ck for 12 5 11 4 V To 12 6 V If an incorrect voltage is present remove cable from the jack and check open circuit voltage at the end of the power supply cable connector If the voltage is still incorre...

Page 15: ...locked If the infant becomes too warm verify that the fan is functioning correctly To reset Indicator LED turn device off and on again Device not used within specified operating temperature Make sure...

Page 16: ...000 operational hours If the maximum adjustment has been reached replace LED PCB as part of routine maintenance Contact Natus Customer Service or authorized service provider to purchase a replacement...

Page 17: ...ot be reached replace LED PCB see Chapter 4 Contact Natus Technical Service or authorized service provider Internal wires are in contact with the fan Open device and verify that the wires do not come...

Page 18: ...Chapter 2 after any repair is completed and the UUR is reassembled The UUR can be disassembled down to all major component parts including Removing the Plastic Cover Section 4 1 Removing the LED PCB S...

Page 19: ...further repair is required return to Chapter 2 and complete Performance Verification and Safety Tests If further disassembly is required store the plastic cover in a protected area and proceed to Sec...

Page 20: ...pleted before beginning these steps 1 The switch is held in the light enclosure by spring tabs two on each side of the switch See Figure 7 Figure 7 Switch Removal 2 Press the tabs against the switch b...

Page 21: ...se all previous steps in chapter 4 return to Chapter 2 and complete Performance Verification and Safety Tests 4 5 About the Device Timer The neoBLUE cozy light is equipped with a timer to track the to...

Page 22: ...ER SUPPLY NOT SHOWN 9 800248 SCREW 18 8 SS PAN HEAD SCREW M4 8MM 10 800247 SCREW 18 8 SS FLAT HEAD SCREW M4 10MM 11 400187 LED LENS 12 400188 LED INTERCONNECT CABLE 13 500044 LED RED 14 030760 DISPOSA...

Page 23: ...eakage current 100 A Earth leakage current 250 A Audible Noise 60 dB Dimensions Width x Length x Height Enclosure 12 in 30 5 cm x 25 5 in 64 8 cm x 4 0 in 10 2 cm Weight 11 0 lbs 5 0 kg Enclosure Powe...

Page 24: ...ed to a greater intensity delivering a higher dosage treatment Please refer to the Section 2 2 2 Adjusting Light Intensity The light intensity output of the neoBLUE cozy device is measured with the ne...

Reviews: