Guidance and manufacturer's declaration - electromagnetic immunity - for equipment and systems within
Professional Healthcare use environment
HI-PRO 2 is intended for use in the electromagnetic environment specified below. The user of HI-PRO 2
should ensure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment -
guidance
Conducted RF
IEC 61000-4-6
3 V rms
150 kHz to 80 MHz
6 V rms
ISM
Bands andAmateur
3 V rms
150 kHz to 80 MHz
6 V rms
ISM
Bands andAmateur
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to 2.7 GHz
10 V/m
80 MHz to 2.7 GHz
Proximity fields from
RF wireless com-
munications
IEC 61000-4-3
27 V/m
386 MHz
28 v/m
450 MHz,
9 v/m
710 MHz, 745 MHz, 780
MHz
28 v/m
810 MHz, 870 MHz, 930
MHz,
28 v/m
1720 MHz, 1845 MHz,
1970 MHz
28 v/m
2450 MHz,
9 v/m
5240 MHz, 5500 MHz,
5785 MHz
27 V/m
386 MHz
28 v/m
450 MHz,
9 v/m
710 MHz, 745 MHz, 780
MHz
28 v/m
810 MHz, 870 MHz, 930
MHz,
28 v/m
1720 MHz, 1845 MHz,
1970 MHz
28 v/m
2450 MHz,
9 v/m
5240 MHz, 5500 MHz,
5785 MHz
Separation distance between any
electronic parts of HI-PRO 2 and
any RF wireless communication
equipment must be more than 30
cm (11.8 inches).
Note
: These guidelines may not
apply in all situations. Elec-
tromagnetic propagation is
affected by absorption and reflec-
tion from structures, objects and
people.