Why do you need to do a cleaning job after each 

measurement?
Because infrared temperature 

uses a

 highly sensitive 

technique to detect the temperature of the target object, 
not only will any earwax 

or

 dust affect the measurement 

accuracy, 

but it is 

more likely to cause 

bacteria 

infection. So we suggest that you cleaning 

as

 well as 

the 

diagram

 shows after 

each use.

12

Frequently Asked Questions and Solutions

1.

(When the measurement probe 

is not properly placed in the ear 
canal or if the measurement 
distance is too far during the 
measurement process, the 
measurement result may be low.)

2.

When measurement probe is 

dirty, the measurement value may 
be low; 

one

 should use a cotton 

swab 

dipped in

 alcohol to wash 

the measurement probe properly.

Screen 
display

Reasons

Solutions

HI

When the target object 
temperature is higher than 
measurement range 

(

the ear 

temperature is more than 

109

.

4°F)

the LCD screen 

will show "HI".

Lo

When the target object 
temperature is lower than 
measurement range

 (t

he 

ear temperature is less 
than 

93.2°F)

, the LCD 

screen will show "Lo"

.

Er.H

This product has 

a max

 

operating temperature of 

104

°

F

. When the 

temperature exceeds this  
point, the LCD screen will 
display an error message 
“Er.H".

When you operate this product, 
the environment temperature 
cannot be greater than 

104°F

.

Er.L

This product has a 

lower

 

operating temperature of 

5

F

. When the ambient 

temperature exceeds this 
temperature point, the LCD 
screen will display an error 
message “Er.L”.

When you operate this product, 
the environment temperature 
can not be lower than 

5

0

°

F

The measurement

 temperature is low.

The measurement 
position is not 
correct.

There

's a blockage

 

in the sensor or ear 
canal.

Please clear 

all

 

debris

 before  

measurement.

Measure the 
temperature correctly 
according to the 
instructions.

Big temperature 
fluctuations with 
continuous 
measurement.

The measurement 
interval is too small.

The interval for each 
measurement should 
be above 10 seconds.

Troubleshooting

Phenomenons

Reasons

Solutions

When the power 
is turned on, the 
screen can not 
be displayed.

The battery is 
exhausted

.

Replace 

with a

 new 

battery

.

The battery 
polarity is wrong.

The battery polarity 
is the same as the 
battery case.

11

When the operation is wrong, if the sound is turned on, then the sound will  
read the value and remind at this moment.(A short beep will be heard.)

Err

When the environment 
temperature changes rapidly 
about 5 degrees, it will show
“ Err ” during measurement 
in object temperature mode, 
then shutdown automatically.

When it shows “Err”,  please place 
this product and keep steady in 
current environment for 30 minutes 
before next measurement.

Summary of Contents for PG-IRT1603

Page 1: ...mometer Thanks for buying and using this product please read this manual carefully before use User manual The USB Lighter Company LLC Address 10940 Wilshire Blvd STE 2010 Los Angeles CA 90024 949 682 8817 Brand NCT Date 06 16 2020 Rev A 0 ...

Page 2: ...ease do not use this product for purposes other than body temperature measurement Do not use mobile phones close to this product Please do not use equipment that generates electromagnetic fields near the product Please do not disassemble or repair this device by yourself Please do not bend or stretch the device Please do not impact or drop this product Caution please refer to attached file Legend ...

Page 3: ...outh When the infrared sensor is damaged or dirty it may cause abnormal measurement results 3 If there is a temperature difference between the storage site and the measurement site please place it at room temperature measurement site for about 30 minutes before next measurement May result in incorrect measurement results 4 Please keep this product out of the reach of children Children can potentia...

Page 4: ...idue of this product at the end of the service life according to local laws and regulations Common sense about body temperature The comparison of different measurement methods The measured values are different if we use different measurement methods The WHO provides normal human body temperature reference values please see below table about the specific temperature difference The changes in human ...

Page 5: ...s low LCD shows sign and battery mark is always on which means you should replace the battery Warm prompts If storing the product for an extended period please remove the battery to assure longevity Battery fluids may harm the product and pollute our environment It is recommended to use alkaline batteries The way to handle a scrapped battery should match the requirements of local country governmen...

Page 6: ... Measurement result Note If you did not hear the beep sound this means that the temperature measurement has not yet been completed Please do not remove the thermometer probe from forehead at this time If you have disabled the sound indication it will have no sound indication 3 Object pattern measurement The installation of the head cover of forehead temperature has no influence to object pattern m...

Page 7: ...H When you operate this product the environment temperature cannot be greater than 104 F Er L This product has a lower operating temperature of 50 F When the ambient temperature exceeds this temperature point the LCD screen will display an error message Er L When you operate this product the environment temperature can not be lower than 50 F The measurement temperature is low The measurement posit...

Page 8: ...kV 4 kV 8 kV 15 KV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 30A m 50 60Hz Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment Specifications of the product Product name Infrared Forehead Ear Thermometer Model number...

Page 9: ...odel PG IRT1603 Infrared Thermometer is used exceeds the applicable RF compliance level above the Model PG IRT1603 Infrared Thermometer should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as re orienting or relocating the Model PG IRT1603 Infrared Thermometer d Over the frequency range 150 kHz to 80 MHz field strengths should ...

Page 10: ...ed Maximum output of transmitter W Separation distance according to frequency of transmitter 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 7 GHz P V d 5 3 1 P E d 5 3 1 P E d 7 1 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be estimated usin...

Reviews: