31
Profiler
User Manual
3.4.2 Contrast (CRTs only)
Open the measurement dialog by clicking <Measure>. The contrast of
your monitor should initially be set to maxi- mum. Position the instru-
ment on the measurement patch and start the measurement by clicking
<Measure>. Please adjust the contrast via the OSD until the indicator is
in the center of the diagram. When you are finished, click <Done> and
<Close> to go back to the main window.
3.4.3 Color temperature (LCDs and CRTs)
Most of the monitor OSDs offer various possibilities to set the color
temperature. If possible select a mode, that allows you to adjust each
channel (R, G and B) separately. This is the most precise method to
preset your monitor to the desired color temperature.
If this option is not available, please select a specific color temperature
(in Kelvin) from the OSD. If the measured color temperature should not
match, preferably select one that is too high than too low.
After clicking
<Measure>
the measurement window will show up. Position
the instrument on the screen and start the measurement by clicking the
<Measure>
button..