38 NED
XCM4040SAT4 UME-0002-04
4.8.3
External Trigger Exposure Mode (Trigger Level)
In external trigger exposure mode
(Trigger Level)
, the exposure time is
determined by the setting for the line period parameter, each exposure starts
with the rising edge and the line period is determined by high trigger pulse
time. The range of programmable exposure time, the timing chart of the
exposure and the readout are shown below.
Table 4-8-3-1 Programmable Exposure Time
4040SAT4
r
Readout time
25.6
a
Trigger pulse H time
≧
24.4
b
Trigger pulse L time
≧
2.4
c
Trigger pulse cycle
≧
26.8
(unit:μs)
Exprosure
Readout
(1)
(1)
(2)
Trigger
(2)
(1)
(2)
(3)
(3)
(CC1)
c
b
a
r
a
Figure 4-8-3-1 External Trigger (Trigger Level) Exposure Mode
The data of Exposure (1) is read out at Readout (1)