Performance Verification
N-550
15
18. Wait for POST to be completed. The %SpO
2
and pulse rate alarm limits
will be reset to the default values.
19. Press the Pulse Rate Alarm Limit button two times rapidly. Within three
seconds press and hold the Adjust Up button. Verify that the pulse rate
display cannot be adjusted above the upper alarm limit setting minus one
point.
20. Press the
SatSeconds
Alarm Limit button. Verify that the N-550 emits a
single beep and that the pulse rate display indicates an alarm limit of
“OFF” for three seconds. Verify that “SAS” is displayed in the %SpO
2
display.
Figure 8: Adjusting SatSeconds Alarm Limit
21. Press the
SatSeconds
Alarm Limit button. Within three seconds press the
Adjust Up button or the Adjust Down button. Verify that the pulse rate
display changes between 10, 25, 50, 100, and OFF.
22. Press the Power On/Off button to turn the N-550 off.
23. Press the Power On/Off button to turn the N-550 on.
24. Press and release the %SpO
2
Alarm Limit button. Verify that the %SpO
2
display indicates an alarm limit of “100.”
25. Press the %SpO
2
Alarm Limit button two times rapidly. Verify that the
%SpO2 display indicates an alarm limit of “85.”
26. Press and release the Pulse Rate Alarm limit button. Verify that the pulse
rate display indicates an alarm limit of “170.”
Summary of Contents for OxiMax N-550
Page 1: ...N 550 Pulse Oximeter Service Manual min min SpO2 N 550 SatSeconds SpO2 ...
Page 10: ...Blank Page ...
Page 38: ...Blank Page ...
Page 64: ...Blank Page ...
Page 81: ...Disassembly Guide N 550 71 Figure 39 Equipotential Ground Terminal Removed ...
Page 82: ...Blank Page ...
Page 86: ...Spare Parts 76 Figure 40 Exploded View ...
Page 90: ...Blank Page ...
Page 108: ...Blank Page ...
Page 120: ...Blank Page ...
Page 130: ...Figure 56 Main PCB Parts Locator Diagram 129 Technical Description N 550 ...
Page 131: ...Figure 57 Front Panel Parts Locator Diagram 131 N 550 Technical Description ...
Page 136: ...Blank Page ...
Page 137: ... ...