N-550
29
A u d i b l e A l a r m S e t t i n g s a n d S e r v i c e
F u n c t i o n s
Introduction
This section discusses use of the service menu to reconfigure power-on
default values, and how to control the behavior of the audible alarm.
Audible Alarm Settings
The following paragraphs describe how to change the behavior of the audible
alarm. Operators can select the volume of the alarm and the duration of alarm
silence. Controls for the N-550 are shown in Figure 9.
Figure 9: N-550 Controls
Alarm Silence State
Press the Alarm Silence button to silence the alarm. Press the Alarm Silence
button a second time to turn the alarm back on.
1 — Power On/Off Button
5 — Pulse Rate Alarm Limit Button
2 — Alarm Silence Button
6 —
SatSeconds
Alarm Limit Button
3 — Adjust Up Button
7 — SpO
2
Alarm Limit Button
4 — Adjust Down Button
Summary of Contents for OxiMax N-550
Page 1: ...N 550 Pulse Oximeter Service Manual min min SpO2 N 550 SatSeconds SpO2 ...
Page 10: ...Blank Page ...
Page 38: ...Blank Page ...
Page 64: ...Blank Page ...
Page 81: ...Disassembly Guide N 550 71 Figure 39 Equipotential Ground Terminal Removed ...
Page 82: ...Blank Page ...
Page 86: ...Spare Parts 76 Figure 40 Exploded View ...
Page 90: ...Blank Page ...
Page 108: ...Blank Page ...
Page 120: ...Blank Page ...
Page 130: ...Figure 56 Main PCB Parts Locator Diagram 129 Technical Description N 550 ...
Page 131: ...Figure 57 Front Panel Parts Locator Diagram 131 N 550 Technical Description ...
Page 136: ...Blank Page ...
Page 137: ... ...