Audible Alarm Settings and Service Functions
N-550
39
Menu Item 11 (Alarm Silence Restrictions)
This menu item is used to change alarm silence behavior. Three options can
be accessed by first pressing the
SatSeconds
Alarm Limit button, then using
the Adjust Up or Adjust Down button to scroll to the desired number.
•
Option 0 = allows the operator to select Alarm Silence Off. There will
be a reminder tone every three minutes.
•
Option 1 = allows the operator to select Alarm Silence Off. There will
be no reminder tone.
•
Option 2 = does not allow the operator to select Alarm Silence Off.
When the desired option is indicated in the %SpO
2
display, press the
SatSeconds
Alarm Limit button to set the current selection.
Note
:
The low battery audible alarm cannot be disabled.
Pulse Rate Lower
Alarm Limit
30 bpm to Upper Alarm Limit
minus 1
40 bpm
Alarm Silence
Reminder
On or Off
On
Alarm Silence
Duration
Off, 30, 60, 90, 120 seconds
60 seconds
Alarm Silence
Restrictions
•
None
•
Sound reminder
•
Do not allow alarms off
None
Alarm Sound
Selector
1, 2, 3
2 (IAW
IEC60601-1-8)
Alarm Volume
1 to 10
4
Table 4: Factory Default Settings (Continued)
Parameter
Range
Factory
Default
Setting
Summary of Contents for OxiMax N-550
Page 1: ...N 550 Pulse Oximeter Service Manual min min SpO2 N 550 SatSeconds SpO2 ...
Page 10: ...Blank Page ...
Page 38: ...Blank Page ...
Page 64: ...Blank Page ...
Page 81: ...Disassembly Guide N 550 71 Figure 39 Equipotential Ground Terminal Removed ...
Page 82: ...Blank Page ...
Page 86: ...Spare Parts 76 Figure 40 Exploded View ...
Page 90: ...Blank Page ...
Page 108: ...Blank Page ...
Page 120: ...Blank Page ...
Page 130: ...Figure 56 Main PCB Parts Locator Diagram 129 Technical Description N 550 ...
Page 131: ...Figure 57 Front Panel Parts Locator Diagram 131 N 550 Technical Description ...
Page 136: ...Blank Page ...
Page 137: ... ...