Specifications
82
Electrical
Environmental Conditions
Instrument
Power Requirements
100 to 240 volts AC, 18 to 26 volt/amps to be
compliant with IEC 60601
-
1 sub
-
clause 10.2.2
Fuses
qty 2, 2 A, 250 volts, slow
-
blow, IEC (5 x 20 mm)
Battery
The battery provides at least two hours of battery life when new and fully charged
with no alarms, no serial data, while using a pulse simulator set for 200 bpm, high
light and low modulation.
Type
Lead acid
Voltage
12 Volts DC, 1.2 AH
Recharge
11 hours with N-550 turned off
12 hours with N-550 operating
Shelf Life
2 months, new, fully charged battery
After 2 months storage the N-550 will
run for 50% of stated battery life
Complies With
91/157/EEC
Sensors
Wavelength
The wavelength range of the light emitted are near 660 nm to
890 nm.
Operation
Temperature
10 ºC to 45 ºC (50 ºF to 113 ºF)