Routine Maintenance
6
2. Perform the electrical safety tests detailed in
on page 27. If
the N-550 fails these electrical safety tests, refer to
3. Inspect the fuses for proper value and rating (F1 and F2 = 2 amp,
250 volts).
Battery
Nellcor recommends replacing the N-550's battery every two years. When the
N-550 is going to be stored for two months or more, remove the battery prior
to storage. To replace or remove the battery, refer to
If the N-550 has been stored for more than 30 days, charge the battery as
described in
on page 8. A fully discharged battery requires
11 hours with the N-550 in standby, or 12 hours if it is in use, to receive a full
charge. The battery is being charged whenever the N-550 is plugged into an
AC power source.