Performance Verification
18
3. Connect a Nellcor
Durasensor
oxygen transducer (sensor), model
DS-100A, to the sensor cable.
4. Clip the DS-100A to the subject as recommended in the sensor's
directions for use.
5. Press the Power On/Off button to turn the N-550 on and verify that the
N-550 is operating.
6. The N-550 should stabilize on the subject's physiological signal in about
15 to 30 seconds. Verify that the oxygen saturation and pulse rate values
are reasonable for the subject.
Alarm Volume Control
1. Press and hold the Alarm Silence button for more than three seconds.
Verify the following:
•
“Alarm Silence Duration Time” is displayed for approximately three
seconds.
•
After three seconds:
•
a steady tone is heard at the default volume setting
•
the %SpO2 display indicates “VOL”
•
the pulse rate display indicates the default setting of “4”
2. Press the Adjust Down button, while holding the Alarm Silence button,
until an alarm volume setting of “1” is displayed. Verify that the alarm
volume has decreased but is still audible.
Summary of Contents for OxiMax N-550
Page 1: ...N 550 Pulse Oximeter Service Manual min min SpO2 N 550 SatSeconds SpO2 ...
Page 10: ...Blank Page ...
Page 38: ...Blank Page ...
Page 64: ...Blank Page ...
Page 81: ...Disassembly Guide N 550 71 Figure 39 Equipotential Ground Terminal Removed ...
Page 82: ...Blank Page ...
Page 86: ...Spare Parts 76 Figure 40 Exploded View ...
Page 90: ...Blank Page ...
Page 108: ...Blank Page ...
Page 120: ...Blank Page ...
Page 130: ...Figure 56 Main PCB Parts Locator Diagram 129 Technical Description N 550 ...
Page 131: ...Figure 57 Front Panel Parts Locator Diagram 131 N 550 Technical Description ...
Page 136: ...Blank Page ...
Page 137: ... ...