Troubleshooting
144
10030881 Rev. B 12-2008
Service Manual
Error Codes
When the OxiMax N-600x pulse oximeter detects an error condition, it displays “EEE”
followed by an error code of up to three digits.
Figure 120.
Error Condition Screen, Battery Failure
READING
TRENDS...
None
Yes
Yes
When the oximeter
needs more than
four to six seconds to
retrieve trend data
from memory.
Completely retrieve sensor
data or select ABORT.
REPLACE
SENSOR
None
No
Yes
When the sensor or
cable wiring is
defective.
Temporarily acknowledge by
pressing the ALARM SILENCE
key. Replace sensor.
RESET
DEFAULTS?
10
No
Yes
When attempting to
reset oximeter to
factory default
setting by pressing
the RESET softkey on
the PARAMS menu.
Leave default settings as they
are by selecting NO, or revert
back to the original factory
default settings by selecting
YES.
SAVE
DEFAULTS?
10
No
Yes
When attempting to
save current settings
as power-on defaults
by pressing SAVE in
the PARAMS menu.
Confirm the new institutional
default settings by selecting
YES, or leave default settings
as they are by selecting NO,
SENSOR
DISCONNECTED
None
No
Yes
When the sensor is
disconnected from
the pulse oximetry
cable, the cable is not
connected to the
oximeter, or the cable
wiring is defective.
Temporarily acknowledge by
pressing the ALARM SILENCE
key. Connect the cable and
sensor to the oximeter,
reconnect the sensor, or
check all connections. If this
does not clear the condition,
replace the cable and/or
sensor.
SENSOR TYPE
5
No
No
First message
displayed when a
sensor is connected
to the oximeter.
Automatically clears on
time-out.
Table 24.
Common Prompts and Error Messages
Message
Time-
out in
secs
Exit on
Alarm?
Exit on
Alarm
Silence?
Displayed
Resolution
Summary of Contents for OXIMAX N-600X Series
Page 8: ...viii 10030881 Rev B 12 2008 Service Manual...
Page 12: ...Safety Information 12 10030881 Rev B 12 2008 Service Manual...
Page 44: ...Theory of Operations 44 10030881 Rev B 12 2008 Service Manual...
Page 96: ...Managing the Data Port 96 10030881 Rev B 12 2008 Service Manual...
Page 188: ...Oximeter Schematics 188 10030881 Rev B 12 2008 Service Manual Main PCB Schematic Sheet 1 of 13...
Page 189: ...Service Manual 10030881 Rev B 12 2008 189 Main PCB Schematic Sheet 2 of 13...
Page 191: ...Service Manual 10030881 Rev B 12 2008 191 Main PCB Schematic Sheet 4 of 13...
Page 192: ...Oximeter Schematics 192 10030881 Rev B 12 2008 Service Manual Main PCB Schematic Sheet 5 of 13...
Page 194: ...Oximeter Schematics 194 10030881 Rev B 12 2008 Service Manual Main PCB Schematic Sheet 7 of 13...
Page 195: ...Service Manual 10030881 Rev B 12 2008 195 Main PCB Schematic Sheet 8 of 13...
Page 197: ...Service Manual 10030881 Rev B 12 2008 197 Main PCB Schematic Sheet 10 of 13...
Page 200: ...Service Manual 10030881 Rev B 12 2008 200 Main PCB Schematic Sheet 13 of 13...
Page 201: ...Service Manual 10030881 Rev B 12 2008 201 Main PCB Assembly Drawing Front View...
Page 204: ...Oximeter Schematics 204 10030881 Rev B 12 2008 Service Manual...
Page 209: ......