General Introduction
22
10030881 Rev. B 12-2008
Service Manual
Oximeter Features
OxiMax SPD™ Alert Feature
The OxiMax SPD™ Alert (SPD) feature detects patterns of desaturation in adults that
are indicative of repetitive reductions in airflow through a patient's upper airway into
the lungs. Relative reductions in a patient's minute ventilation over a period of time
may cause a progressive drop in alveolar partial pressure of oxygen, leading to
arterial desaturation. If these decreases in ventilation are repetitive, they generate
distinct patterns in the saturation trend. Patterns of repetitive desaturation often
develop gradually over time, increasing in severity. Detection of patterns indicates
that a patient might be suffering progressively severe decrements in airflow that
may increase in acuity if left untreated.
Patterns of desaturation are multiple, sequential occurrences of a desaturation
followed by a resaturation. The SPD feature qualifies patterns of desaturation
resulting from such repetitive reductions in airflow based on specific characteristics.
•
The severity of the desaturation event (the depth of the desaturation during the event)
and the extent of the following resaturation
•
The regularity of the desaturation events (how often the pattern repeats)
•
The slope of the desaturation/resaturation trends that form the events
The SPD feature qualifies these patterns of desaturation over a period of six (6)
minutes. Depending on the sensitivity setting for SPD, patterns that persist may
result in an SPD alarm, alerting the caregiver to the condition.
Figure 4.
Clinically Significant Desaturation Patterns
Reference the
N-600x Operator’s Manual
for details.
SatSeconds™ Alarm Management Feature
The oximeter monitors the percentage of hemoglobin binding sites saturated with
oxygen in the blood. With traditional alarm management, upper and lower alarm
limits are set to alarm at specific SpO
2
levels. When the SpO
2
level
fluctuates near an
alarm limit, the alarm sounds each time it violates the alarm threshold. SatSeconds
To Purchase, Visit
Summary of Contents for OXIMAX N-600X Series
Page 8: ...viii 10030881 Rev B 12 2008 Service Manual...
Page 12: ...Safety Information 12 10030881 Rev B 12 2008 Service Manual...
Page 44: ...Theory of Operations 44 10030881 Rev B 12 2008 Service Manual...
Page 96: ...Managing the Data Port 96 10030881 Rev B 12 2008 Service Manual...
Page 188: ...Oximeter Schematics 188 10030881 Rev B 12 2008 Service Manual Main PCB Schematic Sheet 1 of 13...
Page 189: ...Service Manual 10030881 Rev B 12 2008 189 Main PCB Schematic Sheet 2 of 13...
Page 191: ...Service Manual 10030881 Rev B 12 2008 191 Main PCB Schematic Sheet 4 of 13...
Page 192: ...Oximeter Schematics 192 10030881 Rev B 12 2008 Service Manual Main PCB Schematic Sheet 5 of 13...
Page 194: ...Oximeter Schematics 194 10030881 Rev B 12 2008 Service Manual Main PCB Schematic Sheet 7 of 13...
Page 195: ...Service Manual 10030881 Rev B 12 2008 195 Main PCB Schematic Sheet 8 of 13...
Page 197: ...Service Manual 10030881 Rev B 12 2008 197 Main PCB Schematic Sheet 10 of 13...
Page 200: ...Service Manual 10030881 Rev B 12 2008 200 Main PCB Schematic Sheet 13 of 13...
Page 201: ...Service Manual 10030881 Rev B 12 2008 201 Main PCB Assembly Drawing Front View...
Page 204: ...Oximeter Schematics 204 10030881 Rev B 12 2008 Service Manual...
Page 209: ......