Powering the Oximeter
Service Manual
10030881 Rev. B 12-2008
57
Powering the Oximeter
Power Prerequisites
Caution
If any indicator or display element does not light when the pulse oximeter is turned on, do
not use the oximeter. Instead, contact qualified service personnel, your local Nellcor
representative, or Nellcor’s Technical Services Department.
Caution
During POST (immediately after power-up), confirm that all indicators light, all display
segments turn on, and the pulse oximeter speaker sounds a sequence of three ascending
tones. After the POST process completes, verify that a single one-second tone sounds.
Before using the oximeter in a clinical setting, verify the oximeter is safe and working
properly. Verify proper working condition at each power up by following the
procedure below.
Power-On Self-Test (POST)
At power on, the OxiMax N-600x pulse oximeter performs a power-on self-test
(POST), which tests the oximeter circuitry and functions, then proceeds to the
default display. With a sensor cable and an OxiMax pulse oximetry sensor attached,
the unit is ready to register and record patient trend data.
Power-up performance tests verify both power-on self-test (POST) and power-on
defaults and alarm range limits. For defaults and alarm range, see
Note:
Physiological conditions, medical procedures, or external agents such as dysfunctional
hemoglobin, arterial dyes, low perfusion, dark pigment, and externally applied coloring
agents such as nail polish, dye, or pigmented cream may interfere with the oximeter ability
to detect and display measurements.
To power up the oximeter
WARNING
If you do not hear the power-on self-test (POST) pass tone, do not use the oximeter.
Instead, contact Nellcor’s Technical Services or your local Nellcor representative.
WARNING
Ensure the speaker is clear of any obstructions. Failure to do so could result in an
inaudible alarm tone.
Note:
In addition to serving as the POST pass verification, the POST pass tone also functions as an
audible confirmation that the speaker is performing properly. If the speaker does not
function, the alarm warning sounds cannot be heard.
Summary of Contents for OXIMAX N-600X Series
Page 8: ...viii 10030881 Rev B 12 2008 Service Manual...
Page 12: ...Safety Information 12 10030881 Rev B 12 2008 Service Manual...
Page 44: ...Theory of Operations 44 10030881 Rev B 12 2008 Service Manual...
Page 96: ...Managing the Data Port 96 10030881 Rev B 12 2008 Service Manual...
Page 188: ...Oximeter Schematics 188 10030881 Rev B 12 2008 Service Manual Main PCB Schematic Sheet 1 of 13...
Page 189: ...Service Manual 10030881 Rev B 12 2008 189 Main PCB Schematic Sheet 2 of 13...
Page 191: ...Service Manual 10030881 Rev B 12 2008 191 Main PCB Schematic Sheet 4 of 13...
Page 192: ...Oximeter Schematics 192 10030881 Rev B 12 2008 Service Manual Main PCB Schematic Sheet 5 of 13...
Page 194: ...Oximeter Schematics 194 10030881 Rev B 12 2008 Service Manual Main PCB Schematic Sheet 7 of 13...
Page 195: ...Service Manual 10030881 Rev B 12 2008 195 Main PCB Schematic Sheet 8 of 13...
Page 197: ...Service Manual 10030881 Rev B 12 2008 197 Main PCB Schematic Sheet 10 of 13...
Page 200: ...Service Manual 10030881 Rev B 12 2008 200 Main PCB Schematic Sheet 13 of 13...
Page 201: ...Service Manual 10030881 Rev B 12 2008 201 Main PCB Assembly Drawing Front View...
Page 204: ...Oximeter Schematics 204 10030881 Rev B 12 2008 Service Manual...
Page 209: ......