Adjusting the Volume of Audible Tones
Service Manual
10030881 Rev. B 12-2008
63
2. While continuing to press the ALARM SILENCE key, press and hold the ADJUST UP
or ADJUST DOWN key to increase or decrease the volume.
Note:
Default alarm volume is seven (7). Select the appropriate value for the situation. The alarm
volume adjusts up to ten (10) and down to one (1).
Managing Audible Alarms
To set the alarm silence duration
The Alarm Silence Duration display enables you to adjust the alarm silence duration.
1. With the oximeter in the normal monitoring mode, press the ALARM SILENCE key
until the alarm silence duration setting displays. The alarm silence durations
available are 30-, 60-, 90-, and 120-second intervals.
Figure 23.
Alarm Silence Duration Screen
2. Press and hold the ALARM SILENCE key and the ADJUST UP key below the ALARM
SILENCE key on the oximeter panel to increase the alarm silence duration setting.
3. Press and hold the ALARM SILENCE key and the ADJUST DOWN key to decrease
the alarm silence duration setting.
Note:
Releasing the ADJUST UP or ADJUST DOWN key sets the alarm silence duration.
Note:
Default alarm silence duration is 60 seconds. Select the appropriate value for the situation.
The alarm silence duration adjusts up to 120 seconds and down to 30 seconds.
To disable audible alarms
WARNING
Do not disable the audible alarm function or decrease the audible alarm volume if the
patient’s safety could be compromised.
WARNING
Pressing ALARM SILENCE will keep ALL alarms from sounding for the alarm silence
duration period.
Note:
The ability to set the alarm silence duration to OFF can be enabled or disabled by qualified
service personnel.
Summary of Contents for OXIMAX N-600X Series
Page 8: ...viii 10030881 Rev B 12 2008 Service Manual...
Page 12: ...Safety Information 12 10030881 Rev B 12 2008 Service Manual...
Page 44: ...Theory of Operations 44 10030881 Rev B 12 2008 Service Manual...
Page 96: ...Managing the Data Port 96 10030881 Rev B 12 2008 Service Manual...
Page 188: ...Oximeter Schematics 188 10030881 Rev B 12 2008 Service Manual Main PCB Schematic Sheet 1 of 13...
Page 189: ...Service Manual 10030881 Rev B 12 2008 189 Main PCB Schematic Sheet 2 of 13...
Page 191: ...Service Manual 10030881 Rev B 12 2008 191 Main PCB Schematic Sheet 4 of 13...
Page 192: ...Oximeter Schematics 192 10030881 Rev B 12 2008 Service Manual Main PCB Schematic Sheet 5 of 13...
Page 194: ...Oximeter Schematics 194 10030881 Rev B 12 2008 Service Manual Main PCB Schematic Sheet 7 of 13...
Page 195: ...Service Manual 10030881 Rev B 12 2008 195 Main PCB Schematic Sheet 8 of 13...
Page 197: ...Service Manual 10030881 Rev B 12 2008 197 Main PCB Schematic Sheet 10 of 13...
Page 200: ...Service Manual 10030881 Rev B 12 2008 200 Main PCB Schematic Sheet 13 of 13...
Page 201: ...Service Manual 10030881 Rev B 12 2008 201 Main PCB Assembly Drawing Front View...
Page 204: ...Oximeter Schematics 204 10030881 Rev B 12 2008 Service Manual...
Page 209: ......