Managing the Data Port
84
10030881 Rev. B 12-2008
Service Manual
2. Verify the communications format is compatible with the data port of the
oximeter. If the communications format is compatible, real-time data begins to be
displayed on the computer.
3. Capture the text to a file.
4. Press Ctrl-C to halt the data flow.
5. Import the data file into the spreadsheet. The data can now be manipulated by
the commands of the spreadsheet. Some formatting of the data may be
necessary.
Configuring the Data Port Interface
Obtain printouts or patient data by connecting an appropriate RS-232 cable to the
data port on the back of the OxiMax N-600x pulse oximeter and communicating with
a Nellcor Oxinet™ III monitoring system or personal computer. The data port also
provides analog signals representing SpO
2
, pulse rate, and pulse amplitude, as well
as providing a remote nurse call function.
The OxiMax N-600x pulse oximeter provides a bedside oximeter communication for
interfacing with Philips monitors.
Adjust data port options using the appropriate softkey menu items.
Communication Protocol
Overview
WARNING
Do not silence or decrease the volume of the OxiMax N-600x pulse oximeter audible
alarm if patient safety could be compromised.
Note:
Setting the communication protocol automatically sets the communication baud rate.
Change the baud rate to match the abilities of any attached equipment. Perform the
following procedure to change the baud rate.
To adjust the communication baud rate:
1. Turn on the oximeter by pressing the ON/STANDBY button.
2. Press the SETUP softkey.
3. Press the NEXT softkey twice.
4. Press the COMM softkey.
5. Use the ADJUST UP and ADJUST DOWN buttons to select the desired baud rate.
6. Press the EXIT softkey to set the baud rate. The baud rate setting will be in effect
until the oximeter is powered off.
Summary of Contents for OXIMAX N-600X Series
Page 8: ...viii 10030881 Rev B 12 2008 Service Manual...
Page 12: ...Safety Information 12 10030881 Rev B 12 2008 Service Manual...
Page 44: ...Theory of Operations 44 10030881 Rev B 12 2008 Service Manual...
Page 96: ...Managing the Data Port 96 10030881 Rev B 12 2008 Service Manual...
Page 188: ...Oximeter Schematics 188 10030881 Rev B 12 2008 Service Manual Main PCB Schematic Sheet 1 of 13...
Page 189: ...Service Manual 10030881 Rev B 12 2008 189 Main PCB Schematic Sheet 2 of 13...
Page 191: ...Service Manual 10030881 Rev B 12 2008 191 Main PCB Schematic Sheet 4 of 13...
Page 192: ...Oximeter Schematics 192 10030881 Rev B 12 2008 Service Manual Main PCB Schematic Sheet 5 of 13...
Page 194: ...Oximeter Schematics 194 10030881 Rev B 12 2008 Service Manual Main PCB Schematic Sheet 7 of 13...
Page 195: ...Service Manual 10030881 Rev B 12 2008 195 Main PCB Schematic Sheet 8 of 13...
Page 197: ...Service Manual 10030881 Rev B 12 2008 197 Main PCB Schematic Sheet 10 of 13...
Page 200: ...Service Manual 10030881 Rev B 12 2008 200 Main PCB Schematic Sheet 13 of 13...
Page 201: ...Service Manual 10030881 Rev B 12 2008 201 Main PCB Assembly Drawing Front View...
Page 204: ...Oximeter Schematics 204 10030881 Rev B 12 2008 Service Manual...
Page 209: ......