Performance Verification
NPB-40
25
blood oxygen saturation. The test setup procedure sets up Test 2 for
75 percent blood oxygen saturation.
•
Test 3: Modulation
— This procedure simulates an
XI
M
AX
sensor
attached to a patient indicating low and high pulse strength. The test setup
procedure sets up Test 3 for low pulse strength.
•
Test 4: Light
— This procedure simulates an
XI
M
AX
sensor attached to a
patient indicating low and high light level passing through the patient at
the sensor site. The test setup procedure sets up Test 4 for low light level.
Test Setup
Note: A DEC-4 extension cable may be used between the SRC-MAX and the
NPB-40.
1. Connect the SRC
-
MAX tester to the NPB-40 sensor port.
2. Turn on the NPB-40 by pressing the
Power
button.
3. After the NPB-40 completes POST, the NPB-40:
•
Is in SpO
2
alarm
•
Shows an %SpO
2
of 75 (pass criteria is 73 to 77 %SpO
2
inclusive)
•
Shows a pulse rate of 60 (pass criteria is 57 to 63 bpm inclusive)
1 — Sensor Port
Summary of Contents for Oximax NPB-40
Page 1: ......
Page 5: ...Contents NPB 40 iii Power Supply 81 Power Supply Monitoring 81 Index 103...
Page 8: ...Tables vi This Page Intentionally Blank...
Page 18: ...Introduction 10 This Page Intentionally Blank...
Page 22: ...Routine Maintenance 14 This Page Intentionally Blank...
Page 40: ...Performance Verification 32 This Page Intentionally Blank...
Page 50: ...Troubleshooting 42 This Page Intentionally Blank...
Page 66: ...Spare Parts 58 This Page Intentionally Blank...
Page 70: ...Packing for Shipment 62 This Page Intentionally Blank...
Page 82: ...Specifications 74 This Page Intentionally Blank...
Page 90: ...Technical Discussion 82 This Page Intentionally Blank...
Page 92: ...Technical Discussion 84 This Page Intentionally Blank...
Page 94: ...Technical Discussion 86 This Page Intentionally Blank...
Page 96: ...Technical Discussion 88 This Page Intentionally Blank...
Page 98: ...Technical Discussion 90 This Page Intentionally Blank...
Page 100: ...Technical Discussion 92 This Page Intentionally Blank...
Page 102: ...Technical Discussion 94 This Page Intentionally Blank...
Page 104: ...Technical Discussion 96 This Page Intentionally Blank...
Page 106: ...Technical Discussion 98 This Page Intentionally Blank...
Page 108: ...Technical Discussion 100 This Page Intentionally Blank...
Page 110: ...Technical Discussion 102 This Page Intentionally Blank...
Page 113: ......