Administrative Functions
Service Manual
32
Oxinet III
Pagers Menu
To Set Interval Between Initial Page to Primary Pager and Page to
Secondary Pager:
1
Press (or click) the
Pagers
tab to open the
Pagers
menu as shown in Figure 25.
Figure 25: Pagers Menu — Initial
Table 5: Alarms Settings
Description
This is the message that appears on the Central Station
when the alarm occurs (can be edited by a Nellcor Service
technician only)
.
Abbrev
If pagers are assigned to a patient and the
Yes
option in
the
Send Page
column is selected, this is the pager
message that will be sent to the pager (can be edited by a
Nellcor Service technician only).
Level
The settings in this column determines the priority level of
an alarm:
Low
or
High
(can be edited by a Nellcor Service
technician only).
Workstation
Audible
The settings in this column determine if an audible alarm
sounds on the Central Station when this alarm occurs:
Audible
or
Off
. Default setting is
Audible
.
Central Station
Tone
The settings in this column determine the audible tone for
the alarm:
FivePulse
,
ThreePulse
,
Constant Tone
, or
(none)
.
Send Page
The settings in this column determine whether or not this
alarm will trigger an alarm page, if pagers are assigned to
the patient:
Yes
or
No
.
Snapshot
The settings in this column determine if a snapshot will
automatically be taken when this alarm occurs:
Yes
or
No
.
Summary of Contents for Oxinet III
Page 10: ...Introduction Service Manual 4 Oxinet III...
Page 20: ...Installation Service Manual 14 Oxinet III Figure 3 Wired Configuration System Connection...
Page 24: ...Installation Service Manual 18 Oxinet III Figure 4 Wireless Configuration System Connection...
Page 44: ...Administrative Functions Service Manual 38 Oxinet III...
Page 48: ...Maintenance Service Manual 42 Oxinet III...
Page 58: ...Specifications Service Manual 52 Oxinet III...
Page 61: ......