Oxinet III
43
T r o u b l e s h o o t i n g
Troubleshooting List
Table 8 lists troubleshooting tips for some of the problems that might occur while
operating the Oxinet III system.
Some problems might indicate equipment malfunction. Your Oxinet III system
administrator should be contacted to resolve these types of problems. You can also
contact Nellcor’s Technical Services Department at 1.800.635.5267.
Table 8: Troubleshooting
Symptom
Cause
Corrective Action
Message appears:
Please
wait..Maintenance is running.
This should take few minutes
to complete.
The Oxinet III software
performs database
maintenance automatically:
the software temporarily
closes the program,
performs database file
maintenance, and
relaunches the program at
10:05 am every day. This
maintenance procedure
normally takes less than 45
seconds to run.
During
this brief period, the
system does not process
any incoming data,
including alarms.
No intervention is needed.
Poor cursor control on the
Central Station
Cursor out of alignment.
• Connect a keyboard to the Central
Station and restart it.
• Press
F7
key on the keyboard, then
type the Administrative function pass-
word.
• When the
Elo Touchscreen Properties
dialog box opens, press the
Align
but-
ton. The system will guide you through
the alignment process.
• When complete, press the
OK
button
on the
Elo Touchscreen Properties
dialog box to close this function.
Software Activity Indicator
(colored bars in upper left corner
of screen) stop scrolling
Software activity has
stopped.
• Restart the Central Station.
If the problem still exists, contact
Nellcor’s Technical Services
Department at 1.800.635.5267.
Summary of Contents for Oxinet III
Page 10: ...Introduction Service Manual 4 Oxinet III...
Page 20: ...Installation Service Manual 14 Oxinet III Figure 3 Wired Configuration System Connection...
Page 24: ...Installation Service Manual 18 Oxinet III Figure 4 Wireless Configuration System Connection...
Page 44: ...Administrative Functions Service Manual 38 Oxinet III...
Page 48: ...Maintenance Service Manual 42 Oxinet III...
Page 58: ...Specifications Service Manual 52 Oxinet III...
Page 61: ......