Troubleshooting
Service Manual
44
Oxinet III
No display on the Central Station Central Station loss of
power.
• Ensure that the AC power cord is con-
nected to the power input jack on the
back of the base of the Central Station
and the other end is plugged into one
of the battery backed-up AC outlets or
the UPS.
Monitor not connected.
• Ensure all monitor cables are con-
nected.
The UPS has drained its
battery.
• Recharge the battery on the UPS.
• Replace the battery, if necessary.
The UPS is not operating
properly.
• Replace the UPS.
No current data is being
displayed for an active oximeter
or several oximeters
Data from an active
oximeter is not reaching
the Central Station.
• Make sure the cable is securely con-
nected to the oximeter’s data port.
• Check that router/switch has power
Wired configuration:
• Check the connection between the
oximeter and communication server
• Check that the communication server
has power.
• Check the connection between the
communication server and the router/
switch.
Wireless configuration:
• Check that the transmitter has power.
• Check the connection between the
oximeter and transmitter.
• Check that the access point has
power.
No current data is being
displayed for any oximeters in
the Oxinet III system.
Data from all active
oximeters is not reaching
the Central Station.
• Check that the Cat5 cable connecting
the router/switch to the Central Station
is securely connected to any of the
router/switch’s numbered ports.
• Check that the other end of the Cat5
cable is securely connected to the net-
work port connection on the back of
the Central Station.
• Check that the router/switch has
power.
Wired configuration:
• Check that the communication server
has power.
• Check the connection between the
communication server and the router/
switch.
Wireless configuration:
• Check that the access point has
power.
• Check the connection between the
access point and the router/switch.
Table 8: Troubleshooting (Continued)
Symptom
Cause
Corrective Action
Summary of Contents for Oxinet III
Page 10: ...Introduction Service Manual 4 Oxinet III...
Page 20: ...Installation Service Manual 14 Oxinet III Figure 3 Wired Configuration System Connection...
Page 24: ...Installation Service Manual 18 Oxinet III Figure 4 Wireless Configuration System Connection...
Page 44: ...Administrative Functions Service Manual 38 Oxinet III...
Page 48: ...Maintenance Service Manual 42 Oxinet III...
Page 58: ...Specifications Service Manual 52 Oxinet III...
Page 61: ......