IEC 60601-1-1 Compliance
The following information provides instructions and guidance regarding the use
of the Nellcor N-395, N-550, N-595, N-560, and N-600 pulse oximeters. These
oximeters have been found to be compliant with IEC 60601-1-1, when used in a
system level configuration with the AC Y-Cord and the Oxinet III transmitters.
Note that the AC Y-Cord is designed with one AC plug connection to AC mains
power and with two (IEC-320 style) connectors for powering an oximeter and a
transmitter.
The requirements listed below must be followed:
1. This system is authorized for use in the USA only.
2. The AC mains voltages must be in the range of 100 to 132 VAC.
3. The combinations of equipment listed below are authorized to be in the
patient environment.
• N-395 with Oxinet III transmitter, both AC powered by AC Y-Cord
• N-595 with Oxinet III transmitter, both AC powered by AC Y-Cord
• N-550 with Oxinet III transmitter, both AC powered by AC Y-Cord
• N-560 with Oxinet III transmitter, both AC powered by AC Y-Cord
• N-600 with Oxinet III transmitter, both AC powered by AC Y-Cord
Note:
When using Oxinet III, use only the “Lead Year, Model ADP-0502-5V
AC/DC Adapter”. The adapter is connected to mains power via the AC Y-Cord.
WARNING: No other devices are intended for use with the AC Y-Cord. The
use of equipment not specified in this document may subject the patient to
excessive leakage current. The Oxinet III has a DB-9 connector with a metal
shell. Do not touch the metal shell and the patient simultaneously. High
leakage current is possible in case of ground failure of the PSU. During
cleaning, sterilization, disinfection or adjustment of any kind, the
instructions contained in this manual must be followed.
Please contact the Nellcor’s Technical Service Department at
1.800.635.5267
if
you have any questions regarding the proper use of the system.
Summary of Contents for Oxinet III
Page 10: ...Introduction Service Manual 4 Oxinet III...
Page 20: ...Installation Service Manual 14 Oxinet III Figure 3 Wired Configuration System Connection...
Page 24: ...Installation Service Manual 18 Oxinet III Figure 4 Wireless Configuration System Connection...
Page 44: ...Administrative Functions Service Manual 38 Oxinet III...
Page 48: ...Maintenance Service Manual 42 Oxinet III...
Page 58: ...Specifications Service Manual 52 Oxinet III...
Page 61: ......