19
Electric Connection
Electric installations of NeoQi® equipment
should be carried out by qualified specialists
or official representative of NeoQi® AS.
NeoQi® Saltarium is CLASS 1 EQUIPMENT in
which protection against electric shock does
not rely on BASIC INSULATION only, but
which
includes
an
additional
safety
precaution,
which
implies
that
the
EQUIPMENT must be connected to a
protective earth conductor in a fixed wiring,
in such a way that accessible metal parts
cannot become live in the event of a failure
of the basic insulation.
NeoQi® Saltarium belongs to the group of
TYPE B EQUIPMENT, which provides a
degree of protection against electric shock,
which implies:
-allowable leakage current (10 mA);
-reliability
of
the
protective
earth
connection.
NeoQi® Saltarium cannot be used if the air
contains
FLAMMABLE
ANAESTHETIC
MIXTURE WITH AIR or WITH OXYGEN OR
NITROUS OXIDE.