NEULOG PH LOGGER SENSOR GUIDE

 

 

 

NeuLog pH logger sensor 
NUL-206 

 
The pH Sensor can be used for any science experiment or activity where pH readings 
are required some fields of study include Chemistry, Pharmacology, Biochemistry, 
Biology, Earth Science, Environmental Science, etc.   
 
Just a few of the thousands of possible experiments that can be done with the NUL-
206 sensor are: Monitoring of ecological systems, effect of pH on enzymes, static pH 
measurements, chemical reactions with changing pH such as titrations, waste 
measurements, water testing, etc. 
 
The unit of measurement for this sensor is the pH scale. 

 

pH: The universal measurement for pH which directly measures the 
concentration of hydrogen ions using the scale of 0 to 14. 

 

Sensor calibration: 

 
Through usage, the pH probe deviates slightly from its stored voltage value 
corresponding to a pH of 7 (the offset voltage).  The conversion formula must be 
adapted to this change by calibrating the pH sensor to a pH 7 buffer solution.  After 
calibration, the information will be stored for later testing.  For the most accurate 
readings, recalibrate regularly prior to experimentation. 
 
 
 

Materials Needed

:  

 

 

Neulog-206 pH Sensor 

 

pH  7 Buffer Solution 

 

1 of the following (depending on experiment type) 

o

 

USB-200 USB Module (

For On-line experiments)

 

o

 

BAT-200 Battery (

For Off-line experiments)

 

o

 

VIEW-101 Viewing Module and BAT-200 Battery 

(For use with 

Neulog Viewer)

 

 

Procedure: 

 

1. 

The pH Sensor needs a voltage source to properly calibrate so connect the 
pH Sensor to: 

a. 

For On-line experiments

:  USB-200 USB Module. 

b. 

For Off

-line experiments

:  BAT-200 Battery 

c. 

For use with the Neulog Viewer

:  

VIEW-101 Viewing Module 

and BAT-200 Battery. 

2. 

Remove the pH Sensor probe’s tip from the plastic storage solution bottle. 

3. 

Rinse the pH Sensor’s tip with distilled water. 

4. 

Insert pH Sensor’s tip into pH 7 buffer, if no buffer is available you can use 
tap water (although it is not as accurate). 

5. 

The sensor now needs to be calibrated: 

a. 

For On-line experiments

i.  Connect the pH Sensor to the left side of a USB-200 

Module (which is connected to a computer via the USB 
to mini USB cable.) 

ii.  Open the Neulog software 

iii. 

Once the pH Sensor is detected click the “Module 
setup” button. 

iv. 

Click the “Calibration icon” (looks like a scale) to 
automatically calibrate 

b. 

For Off-line experiments and for use with Neulog Viewer:

 

i.  Press an

d hold the pH Sensor’s push button for 3 

seconds 

6. 

The pH Sensor is now calibrated for use. 

 

Quick start procedure: 

 

PC or Mac Computer 

 

Materials needed: 
 

 

NUL-206 PH Sensor 

 

USB-200 USB Module 

 

A USB to mini USB cable (which comes with the USB-200) 

 