LabLynx User Manual Revision 1.1 2/18/2021
16
Start Cheetah and Turn LabLynx On
Open Cheetah using the configuration file you edited above, Cheetah.cfg. While Cheetah
is loading, turn on the LabLynx using the push button on the side. When the LabLynx is
turned on, the status LED will turn blue. After the LabLynx has booted and connected to
Cheetah, the status LED turns green. LED codes are described in
•
Cheetah will attempt to connect with the LabLynx for 60 seconds before it times
out.
•
The LabLynx detects connected headstages during bootup. Therefore, headstages
must be connected to the LabLynx before turning the LabLynx on. Headstages
connected after bootup will not function.
Connect LabLynx to Test Setup
Step
Description
1
Set the Minirator, or other signal source, to output a 1 V
PP
sine wave at 100 Hz.
The Signal Mouse will reduce this signal to approximately to 1 mV
PP
.
2
Connect the Minirator, or other signal source, to the Signal Mouse using a BNC
cable.
3
Connect HS-64-mux to the two Omnetics connectors located on the HS-64-mux
Signal Mouse
4
Turn the Bank 1, 2, 3, and 4 switches on the Signal Mouse to the Signal position
(Up). Turn the Reference switch on the Signal Mouse to the Ground position
(Down).
5
Connect the uDB-25 connector from the HS-64-mux into connector 1 of the
LabLynx.