LabLynx User Manual Revision 1.1 2/18/2021
17
Test the LabLynx with a Test Input Signal
In Cheetah, select the
ACQ
button to start acquisition. Set the Input Range for all 64
CSCs to 1000 µV. Each should show approximately 1 mV
PP
sine waves.
Figure 4-6 Bank 1, 2, 3 and 4 Switches Up. Reference Switch Down
Switch the Bank 1 switch to the Ground position (down). CSCs 1-8 and 33-40 should
now be flat-lined while CSCs 9-32 and 41-64 should still show 1 mV
PP
sine waves.
Figure 4-7 Bank 2, 3 and 4 Switches Up. Bank 1 and Reference Switches Down