31
Innova 4200
•
M1233-0053
User’s Guide
7.10 Gassing Cultures in the Innova 4200
There is an optional gassing manifold (M1233-0291) available for the 4200. The gassing
manifold is installed in the gassing port on the right hand side of the unit. Instructions on
how to install the gassing manifold are included with the manifold.
7.11 Accessory Recorder Connector
To record speed and temperature, an accessory recorder can be installed on the back
panel of the shaker (
see Figure 4
). The recorder should have the following capabilities:
•
Two channels, required for Speed and Temperature. Each channel should have signal
conditioning which accepts 0-5 volt input.
•
The pin-out diagram and scale below identifies the application.
A mating connector is required on the recorder cable (not supplied).
(This is a 9-pin male D subminiature connector—AMP Amplimite HDP-20 series or
equivalent.)
Figure 11: Detail of 9-Pin Recorder Connector
5 4 3 2 1
9 8 7 6
Pin Number
Signal Name
Scale
6
2
Speed
Ground
1V = 100 RPM
7
Temperature
1V = 20°C
3 Ground
Summary of Contents for innOva 4200
Page 2: ...2 New Brunswick Scientific User s Guide...
Page 57: ...57 Innova 4200 M1233 0053 User s Guide Figure 21 Electrical Schematic Quadrant A...
Page 58: ...58 New Brunswick Scientific User s Guide Figure 22 Electrical Schematic Quadrant B...
Page 59: ...59 Innova 4200 M1233 0053 User s Guide Figure 23 Electrical Schematic Quadrant C...
Page 60: ...60 New Brunswick Scientific User s Guide Figure 24 Electrical Schematic Quadrant D...
Page 66: ...66 New Brunswick Scientific User s Guide Figure 30 100V Power Schematic Overview...
Page 67: ...67 Innova 4200 M1233 0053 User s Guide Figure 31 100V Power Schematic Quadrant A...
Page 68: ...68 New Brunswick Scientific User s Guide Figure 32 100V Power Schematic Quadrant B...
Page 69: ...69 Innova 4200 M1233 0053 User s Guide Figure 33 100V Power Schematic Quadrant C...
Page 70: ...70 New Brunswick Scientific User s Guide Figure 34 100V Power Schematic Quadrant D...