47
Figure 22: Program 1 – Step 1
PRG1 - STEP -- 01
Time
00:00
ºC
20.0
RPM
Off
UV
Off
GRO
Off
Save Cancel
Step number
Use these arrows to
scroll through steps.
DO NOT USE until programming
is complete.
Use to leave
Programming mode
without saving any
NOTE:
If your shaker is not equipped with these optional features, “UV” and
“GRO” will appear with the word “Off” in this screen but will not be
programmable.
3. Turn the Knob to highlight the Time setting (00:00 in the sample screen
above), then click the Knob in. The field will flash.
4. Turn the Knob until the desired running
duration
for this step (from 00:01,
which means one minute, to 99:59) appears, then click the Knob in to save the
setting. We will set Step 1 time to eight hours for this example (
see Figure 23
).
5. Turn the Knob to highlight the
ºC
temperature setting (
20.0
in the sample
screen above), then click the Knob in. The field will flash.
6. To set the temperature desired (
ºC
from 4.0 to 80.0) for the time period you
have set, turn the Knob (left to decrease, right to increase). When the desired
value appears, click the Knob in to save the setting. We will set Step 1
temperature to 37.0ºC (
see Figure 23
).
7. Turn the Knob to highlight the
RPM
setting (
Off
in the sample screen above),
then click the Knob in. The field will flash.
8. Turn the Knob to select the desired shaking speed (25 to 400 RPM) for this
time period, then click the Knob in to save the setting. We will set Step 1
speed to 150 RPM (
see Figure 23
).
new settings.
New Brunswick Scientific Co., Inc.
User’s Guide
Summary of Contents for Innova 43
Page 2: ...2 Innova 43 43R Incubator Shakers M1320 0050 User s Guide ...
Page 10: ...10 Innova 43 43R Incubator Shakers M1320 0050 User s Guide ...
Page 79: ...79 11 1 1 CE Declaration of Conformity New Brunswick Scientific Co Inc User s Guide ...
Page 92: ...92 Filling the 33 Wavy Arrows Icon 21 Innova 43 43R Incubator Shakers M1320 0050 User s Guide ...