63
9.8
Replacing Optional Germicidal Lamp
WARNING!
Before installing or removing the germicidal “UV” lamp, the authorized
technician will:
(1) Make sure power is completely disconnected from the unit,
(2) Wear plastic gloves to protect the lamp from skin oils and to protect
your hands just in case the lamp should break,
(3) Before removal, make sure the lamp has fully cooled.
To replace the UV lamp, your service technician will follow all the
WARNINGS
above
and the procedure and
WARNINGS
outlined below, with reference to Figure 29:
1. Make sure power is disconnected from the shaker.
2. Open the Service Compartment by removing the side panel.
3. Free the heater cable harness and the temperature probe from their tie wraps and
Permagum
®
.
Figure 29: Germicidal (UV) Lamp Replacement
NOTE that wires, tie wraps and Permagum
®
are not shown above.
4. Remove the evaporator cover from the Service Compartment, paying careful attention
to the drain tubes.
5. Wearing plastic gloves, remove the germicidal lamp by pressing it toward the spring-
loaded tombstone. The lamp has a single pin at each end.
6. Still wearing plastic gloves, install a new lamp.
7. Pull the heater cable harness and temperature probe back through the holes in the
Service Compartment.
New Brunswick Scientific Co., Inc.
User’s Guide
Summary of Contents for Innova 43
Page 2: ...2 Innova 43 43R Incubator Shakers M1320 0050 User s Guide ...
Page 10: ...10 Innova 43 43R Incubator Shakers M1320 0050 User s Guide ...
Page 79: ...79 11 1 1 CE Declaration of Conformity New Brunswick Scientific Co Inc User s Guide ...
Page 92: ...92 Filling the 33 Wavy Arrows Icon 21 Innova 43 43R Incubator Shakers M1320 0050 User s Guide ...