- 44 -
9-3. Sampling Unit Replacement
1. Remove the sensor unit by taking steps 1 to 4 of 9-2.
“Sensor Unit Installation/Replacement”.
2. While pressing the front panel
’s bottom with both thumbs, slowly press the panel forward to
remove it from the main unit.
3. Disconnect the two connectors from the front panel.
Turn off the gas detector prior to sampling unit replacement. Setting the main unit power
switch to the OFF position will turn off the trouble alarm output (open collector). If the
trouble alarm output is used to operate the interlocks of the external devices, release
the interlocks beforehand, as needed, to prevent a possible activation of the interlocks
during replacement.
Press with thumbs
Disconnect the connector from the front panel.
Disconnect the connector from the pump unit.
Sampling unit Front panel
Press here (2 places)