24
Prior to conducting the procedure determine accurate
direction to North, West, South and East, accuracy
of these references will define accuracy of the
compass calibration.
- Turn on the LRF.
- In the LRF menu select «SET», then «CAL». Press
ACTION.
- Then select «CALS» and press ACTION
- When the display shows «C 0 °», the LRF shall be
set horizontally and aim the LRF to the North (0
degrees), press ACTION.
- When the display shows «C 90 °», the LRF shall
be set horizontally and aim the LRF to the East (90
degrees), press ACTION.
- When the display shows «C 180 °», the LRF shall
be set horizontally and aim the LRF to the South
(180 degrees), press ACTION.
- When the display shows «C 270 °», the LRF shall
be set horizontally and aim the LRF to the West
(270 degrees), press ACTION.
- When the display shows «corr» press MODE to
exit the calibration mode.