Data Recall mode
In
Data Recall
mode results of the last 10 measurements can be
displayed.
Measurement results are saved in on-board memory in sets depending on
the parameters set for measuring, for instance, if only distance is
measured - the set will consist of one number, if distance and azimuth are
measured – the set will consist of two numbers.
To enter
Data Recall
mode select
rEc
from
mode
menu. First
measurement set number (display name:
rEc
1
) will start flashing.
Pressing M button moves the unit along the list of measurement set
numbers and measurement results:
rEc 1
, distance 1, azimuth 1, elevation 1,
rEc 2
, distance 2, azimuth 2, elevation 1, …,
rEc 0
, distance 0, azimuth 0, elevation 0.
Measurement sets are numbered in reverse order: first recalled set is the
result of the last measurement. If any parameter has not been actually
measured it will not be recalled.
Pressing A button at any moment within the list of recalled results brings
the unit to
Ready to Measure
mode.
Choosing
cLr
at the end of recalled data will erase the device
memory.
Computer interconnectivity
The unit supports RS-232 interface. Physical connection is performed
through connector with threaded coupling mechanism, designed for the
heavy-duty outdoor applications.
Note:
Do not apply excessive force when disconnecting cable. Damage
to cable or connector resulting from improper treatment is not covered by
warranty.
To attach the cable to the rangefinder match the key on the plug (1, Fig.
5) with the corresponding key on the socket and then push gently the
plug holding it by its casing (3, Fig. 5) until click.
To detach the cable slide sleeve (2, Fig. 5) away from the unit and pull
the plug out.
The communication port must be initiated prior to connecting the unit to
the computer. The data is transmitted as hex string code with fixed baud
rate of 38400 bit/sec, 8 bits, one stop bit, and no parity.
This connector is IP67 graded.