MCS260B
CORNERSTONE 260B MONOCHROMATORS
39
13 SPECIFICATIONS
General Specifications
Focal Length
260 mm
F/#
F/3.9
Wavelength Selection Method
Motorized
Usable Wavelength Range
250 to 2500 nm, grating dependent
Spectral Resolution
Grating dispersion and slit width dependent
Wavelength Accuracy
±0.50 nm
1
Wavelength Step Resolution (Precision) ±0.0754 nm
1
Stray Light
0.01%
2
Input Ports
1 input port, expandable to two input sources
using optional 77765 Beam Steerer
Output Ports
Axial (all models) and Lateral (dual port models)
Shutter Control
Software, Hand Controller, low-level commands
External Shutter Control
BNC connector; 0 V or short opens shutter
(normally closed)
Shutter minimum exposure time
0.3 s
3
Shutter maximum repetition rate
2.0 Hz
4
Motorized Filter Wheel Compatibility Models USFW-100, 74010, Apex2 filter wheel
Utility Software Requirements
Windows 7 or 10, 32-bit or 64-bit operating
system
TracQ Basic Software Compatible
Yes
74009 Hand Controller Compatible
Yes
Power Requirements
100-240 VAC, 47-63 Hz
Weight
21 lbs. [9.5 kg]
Scan Speed
24 nm/s at 10 nm step rate
1
; 180 nm/s for a
single step
1
Computer Interface
USB (Type B), RS-232 (DB9)
Grating Size
50 mm
Optical Axis Height
3.52” (89.4mm)
Operating Temperature
25.0 ± 3.0 °C
Storage Temperature
-40 to +70 °C, non-condensing
Humidity
<85% relative humidity, non-condensing
1 Using a 1200 Lines/mm grating measured near blaze wavelength of 400 nm.
2 Measured by comparing the peak throughput of a HeNe laser source at 632.8 nm to ±2 nm off peak at 0.5 nm
spectral bandwidth using 1200 Lines/mm grating blazed at 500 nm.
3 Exposure time measured from closed to open to closed shutter.
4 Using the mechanical shutter as a beam chopper will reduce the lifetime of the shutter and it is not commonly
used for this purpose.