3.35
EDH0162En1040 – 06/99
MM4005
DA — Read desired acceleration
Usage
IMM
PGM
MIP
Syntax
xxDApp
Parameters
Description
xx
[int]
— Axis number.
pp
[int]
— Auxiliary parameter.
Range
xx
—
1
to
4.
pp
—
0
or
1
.
Units
xx
— None.
pp
— None.
Defaults
xx
Missing: Error B.
Out of range: Error B.
Floating point: Error A.
pp
Missing: 0.
≥
1: 1.
Description
This command reads the motion acceleration assigned in the axis setup or
redefined through the AC command. This is the acceleration of a trape-
zoidal type motion profile used in point-to-point moves.
On Power-Up, the acceleration defaults to the value preset in the front
panel SETUP menu.
If
pp
is 0 or missing, the actual value of acceleration is reported. If
pp
≥
1,
the maximum allowed value of acceleration is reported.
Returns
xxDAnn
xx
—
Axis number.
nn
—
Acceleration value, in pre-defined units.
Errors
A
—
Unknown message code.
B
—
Incorrect axis number.
S
—
Communication time-out.
Rel. Commands
AC
—
Set acceleration.
Example
2DA |
Read acceleration of axis #2.
2DA10
|
Controller returns an acceleration value of 10.
2PA15 |
Move to absolute position 15.
WT500 |
Wait for 500 ms.
2AC4 |
Set axis #2 acceleration to 4.
2DA
|
Read acceleration of axis #2.
2DA4
|
Controller returns an acceleration value of 4.
2DA1
|
Read maximum acceleration of axis #2.
2DA80
|
Controller returns a value of 80.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com