3.101
EDH0162En1040 – 06/99
MM4005
QW — Save general parameters
Usage
IMM
PGM
MIP
Syntax
QW
Parameters
None.
Description
This command saves in non-volatile memory all general parameters that
have been changed through on-line commands or from within a program.
For a list and description of the general parameters that are stored in non-
volatile memory see the General Setup paragraph of the Local Mode sec-
tion.
NOTE
During the execution of this command, the communication (IEEE / RS232) is broken off.
NOTE
Since this command is equivalent to making changes in SETUP mode, it is valid only when
motor power is turned off. If the command is issued when the motor power is on, the con-
troller will ignore it and report error D.
NOTE
Before saving, make sure that the new set of parameters is correct and safe to use. Also, it
is good practice to verify that the parameter saving procedure worked properly by issuing
the TB or TE command afterwards.
Returns
None.
Errors
A
—
Unknown message code.
D
—
Unauthorized execution.
Rel. Commands
PW
—
Save parameters.
Example
CMM0B19200 |
Set RS-232-C liaison with 19200 Baud.
… |
… |
Verify the new parameter is working fine.
… |
QW
|
Save general parameters non-volatile memory.
TE |
Read error register.
TE@
|
Controller returns a no error code.