3.173
EDH0162En1040 – 06/99
MM4005
XI — Read integral gain factor
Usage
IMM
PGM
MIP
Syntax
xxXI
Parameters
Description
xx
[int]
— Axis number.
Range
xx
—
1
to
4.
Units
xx
— None.
Defaults
xx
Missing: Error B.
Out of range: Error B.
Floating point: Error A.
Description
This command reads the integral gain factor of an axis
NOTE
The command reads the value actually used in the servo loop. If the PID
parameters are modified using the KI command but the digital filter has
not been updated by sending an UF, the command will still read the old
value.
Returns
xxXInn
xx
—
Axis number.
nn
—
Integral gain factor.
Errors
A
—
Unknown message code.
B
—
Incorrect axis number.
S
—
Communication time-out.
Rel. Commands
KI
—
Set integral gain factor.
TF
—
Read filter parameters.
UF
—
Update servo filter.
Example
2XI
|
Read integral gain factor of axis #2.
2XI0.005
|
Controller returns an integral gain factor of 0.005.