6.17
EDH0162En1040 – 06/99
MM4005
Feature Descriptions Tutorial
Asynchronous Acquisition: AQ Command
The AQ command saves the current position of the axes in the TRACE
buffer and generates a synchronizing pulse.
xx
AQ
nn
To record the current position of the axes at the moment desired.
Where:
xx
—
Axis number from 1 to 4.
nn
—
0 or 1
nn = 0
:
Without pulse.
nn = 1
:
With pulse.
Example
GQ0
|
Initialize of global trace buffer.
NT, FA90
|
Initializing trajectory.
CR10, CA5
|
Element 1.
CA350
|
Element 2.
CA5
|
Element 3.
VV5
|
Set trajectory velocity to 5 units/sec.
ET
|
Displacement with generation of pulses.
WN2,
AQ
|
At the beginning of element 2, axis positions are recorded without
synchronization pulse.
WN3,
AQ1
|
At the beginning of element 3, axis positions are recorded with a
synchronization pulse.
6.16